- 6 -
www.bgb-info.com
Figure 3:
Cap removed,
exposing probe tip
brush for cleaning.
Figure 2:
Cap unscrewed
and partially
removed.
8.0 Maintenance
Avoid spilling liquids onto the unit, or it may malfunction. If a liquid is spilled onto the unit, turn
off the power immediately, remove heavy liquids with a dry towel, and let the unit sit until the liquid
dries. Dust and debris can enter the probe tip of the leak detector and, over time, can clog the small-
bore tubing inside the unit. To prevent this, clean the probe tip periodically. To clean the probe tip,
unscrew the cap to expose the brush (Figures 2 and 3). Gently clean the probe using a small brush or
your fingers to remove dust and debris, then replace the cap. Do not use liquids to clean the probe.
Liquids can damage the leak detector if drawn in through the probe.
Information on where to have the unit sent for maintenance or service* is listed at the end of this
document.
9.0 Troubleshooting
*Contact BGB for return instructions for servicing a damaged unit. Additional charges may apply if the warranty has expired or the unit is damaged due to misuse.
Problem
Possible Cause
Suggested Solution
Sensitivity decreased
Probe clogged
Clean the probe tip to remove any debris
Probe line punctured
Visually inspect probe line for holes*
Response decreased
Detector not zeroed
Re-zero detector
LED bar graph stays lit during operation
Detector re-zeroed before unit was purged out
Allow adequate time for detector to purge, then re-zero
Reference gas inlet covered by hand or other object
Remove obstruction
Does not power up
Batteries need to be charged
Charge unit