3
the Slimax curing light. Even supposing if the eye specialist agrees; prudence is strongly recommended, For the
reason that the intensity of the light could cause accidents. Class 2M protective eyewear is suitable for use with
units emitting radiation at wavelengths between 420 and 480 nm, should be worn at all times.
IMPORTANT
• Batteries must be charged prior to use.
• The fiber-optic light guide is provided in a non-sterile condition and must be sterilized prior to
patient contact.
INTRODUCTION
The Slimax LED (Light Emitting Diode) is a cordless pen style curing light used for the polymerization of
light-cured dental material used by dental professionals. It is powered by a rechargeable Lithium-Ion battery
pack. This curing light system from Beyes Dental Canada Inc. generates a powerful output up to 2000 mW/cm
(± 10%). The curing procedure for composite materials is easier and quicker than ever before with the options
of 5, 10, 15 and 20 seconds curing time. Users can either select Full Power, Ramping Power or Pulse. This slim,
light weight curing light is ergonomically designed for user’s comfort and flexibility. It’s solid aluminum body
sets it apart from the mass of plastic competitors. The quality, durability, and ease of applied infection control
techniques make this curing unit a state of the art instrument. The curing light (Slimax-C) comes in four dif
-
ferent colors, Pink, Silver, Blue, and Purple.
ASSEMBLY
The box contains:
a.
1 Curing Light Controller
b.
1 Battery (The battery contains Lithium-Ion cells. A new
battery should be charged for 6 hours prior to its first use.)
c.
1 Tip
d.
1 Battery Charger
e.
1 Power Supply
f.
1 Light Sheild
g.
1 8-mm Curved Light Guide
OPERATION
BATTERY:
The battery pack contains a Lithium ion (Li-ion) battery. When the battery is
low, you will see “LO” on the LED display panel along with slow frequency beeps.
INSTALL THE BATTERY
1. Remove the battery and the Curing Light Controller with LED Display from the box
2. Gently screw the battery into the Curing Light Controller
3. To remove the battery, gently unscrew the battery from the Controller
a.
b.
g.
f.
c.
e.
d.