BEYES Maxso E600P Instructions For Use Manual Download Page 7

7

5-4 Set the maximum speed using the Rotation Adjustment 

Key (+/-).

5-5 Select FWD/REV rotation using the FWD/REV Select key. 

5-6 The micro motor operation is controlled by the air switch/

foot pedal of the delivery unit. 

5-7 Program Function.

You can memorize the setting:

1) set the values for gear ratio, speed

2) Press P1 or P2 key for more than 3 seconds until it 

beeping, the setting is completed.

Push

Digital Speed Display

Increase

Decrease

FWD/REV Lamp

6. Sterilization

Sterilize the motor only.

For sterilization, we recommend the autoclave sterilization 

method.

Sterilization is required the first time you use the unit as well 

as after each patient as noted below.

Autoclaving

1)Turn off the power.

2) Detach the motor from the motor cord.

3) Clean the surface of the motor with a brush (Do not use a 

metal brush), and wipe it with cotton and disinfectant alcohol.

4) Screw the Motor Cap to the Motor. Put the Autoclave Plug 

onto the motor insert.

5) Insert into autoclave ouch. Seal the pouch.

6) Autoclavable up to a maximum of 135°C.

7) Keep the Motor in the autoclavable pouch until you are 

ready to use it to keep it clean.

 Caution

- Do not autoclave (or any other high temperature sterilization) Control Box, AC Adaptor, or Motor 

Cord.

- Do not lubricate the motor.

- Do not wipe or immerse the system in acidic water or acidic solutions.

- Do not sterilize with dirt on the surface, it may cause rust.

Autoclave Plug

Motor Cap

Tighten

Summary of Contents for Maxso E600P

Page 1: ...Electric Micromotor System Instruction for Use Maxso E600P...

Page 2: ...r by holding on to the base of the plug and not the cord To Prevent Electric Shock Use an electrical outlet which is grounded If you feel any abnormality such as excessive vibration heat or noise stop...

Page 3: ...extended periods of time It can cause overheating Do not connect or disconnect the cord until the drive motor has completely stopped Do not connect disconnect the hand piece during operation Verify th...

Page 4: ...Gear Ration Select Key 5 FWD REV Select Key 6 Program 1 Button 7 Program 2 Button 8 Speed Adjusters 1 9 10 2 3 4 6 5 7 8 12 11 13 14 16 15 9 Motor 10 Motor Cord 11 Power Switch 12 AC Adapter Connecto...

Page 5: ...urbine hose connector and tighten completely Caution Do not use any other hose than the ones provided with the unit Do not use contaminated air Blow out the lines before installation Caution Never use...

Page 6: ...requires lubrication E Type micro motor Following lubrication stand the hand piece attachment vertically until the excess solution drains Wipe the hand piece then attach it to the motor Do not connec...

Page 7: ...se the unit as well as after each patient as noted below Autoclaving 1 Turn off the power 2 Detach the motor from the motor cord 3 Clean the surface of the motor with a brush Do not use a metal brush...

Page 8: ...internal cir cuit more than 3 7V Contact your dealer E8 Overheating error High temperature safety system activated due to extended use Wait for a while and let it cool down before using again 7 Mainte...

Page 9: ...limit of air pressure Check the air pressure on the delivery system Do Foot Air calibration Beeps when turning on the switch You are stepping on the foot pedal when turning on the power switch Do not...

Page 10: ...owing 1 Defects damage or other conditions caused in whole or in part by misuse abuse negligence alteration accident freight damage negligent storage tampering or failure to seek and obtain repair or...

Page 11: ...thorized to sell the product at issue within the medical animal health or dental market as the case may be in which purchaser intends to use the product 10 4 Exclusive Remedy Consequential Damages Dis...

Page 12: ...ch Email info beyes ca Web www beyes ca Printed in Canada File ENI019 Rev 3 06 01 21 Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner licen...

Reviews: