Operation Manual

• 

Equipment Classification

- Class II equipment:  

- Type B shock protection:  

- Recommended sterilization: See 7. Sterilization

- Do not use in the presence of flammable liquids or gases.

 

Cautions for safe handling and operation

Read these safety precautions first before using this product. 

These instructions will show you how to operate this product safely and prevent injury to yourself 

or others. They are classified by the degree and/or severity of danger. All cautions regarding safety 

should be observed. 

Classification

Degree and severity of danger

Warning

Provides an instruction where personal injury may occur.

Caution

Provides an instruction where minor to medium injury may occur.

Notice

Provides an instruction that should be observed for safety reasons.

 Warning

- To Prevent Electric Shock: Do not plug in the power cord with wet hands. 

- To Prevent Electric Shock: Prevent water from getting on to the control box.

- To Prevent Electric Shock: Plug in and remove the power by holding on to the base of the plug and 

not the cord.

- To Prevent Electric Shock: Use an electrical outlet which is grounded. 

- If you feel any abnormality such as excessive vibration, heat, or noise, stop using the unit 

 immediately.

 

 Caution

- When operating this system, always consider the safety of the patient. 

- The product is designed for clinical dental use by qualified personnel

- Every connection must be firm. If it is not firmly connected, it may cause an air or water leak, or the 

LED might not turn on. 

- Do not drop the product or cause it to be impacted in any way. This could cause a malfunction and 

damage.

- This system is to be used with ISO 9168, type B or type C hose connector.

- This system is meant to function in temperatures of 0-40°C (32-104°F), humidity at 10-85% RH, 

atmospheric pressure 700-1060hPa and no moisture or condensation in the Control Unit. Operating 

outside of these parameters can cause a malfunction.

- When installing the Control Box or Motor, do not bend or twist the tubing or the cord. No excessive 

force is need to do the installation.

- Do not use contaminated air (by dust or moisture). If the air contains water or dust , it might cause 

a malfunction or overheating.

Summary of Contents for Maxso E600P

Page 1: ...Electric Micromotor System Instruction for Use Maxso E600P...

Page 2: ...r by holding on to the base of the plug and not the cord To Prevent Electric Shock Use an electrical outlet which is grounded If you feel any abnormality such as excessive vibration heat or noise stop...

Page 3: ...extended periods of time It can cause overheating Do not connect or disconnect the cord until the drive motor has completely stopped Do not connect disconnect the hand piece during operation Verify th...

Page 4: ...Gear Ration Select Key 5 FWD REV Select Key 6 Program 1 Button 7 Program 2 Button 8 Speed Adjusters 1 9 10 2 3 4 6 5 7 8 12 11 13 14 16 15 9 Motor 10 Motor Cord 11 Power Switch 12 AC Adapter Connecto...

Page 5: ...urbine hose connector and tighten completely Caution Do not use any other hose than the ones provided with the unit Do not use contaminated air Blow out the lines before installation Caution Never use...

Page 6: ...requires lubrication E Type micro motor Following lubrication stand the hand piece attachment vertically until the excess solution drains Wipe the hand piece then attach it to the motor Do not connec...

Page 7: ...se the unit as well as after each patient as noted below Autoclaving 1 Turn off the power 2 Detach the motor from the motor cord 3 Clean the surface of the motor with a brush Do not use a metal brush...

Page 8: ...internal cir cuit more than 3 7V Contact your dealer E8 Overheating error High temperature safety system activated due to extended use Wait for a while and let it cool down before using again 7 Mainte...

Page 9: ...limit of air pressure Check the air pressure on the delivery system Do Foot Air calibration Beeps when turning on the switch You are stepping on the foot pedal when turning on the power switch Do not...

Page 10: ...owing 1 Defects damage or other conditions caused in whole or in part by misuse abuse negligence alteration accident freight damage negligent storage tampering or failure to seek and obtain repair or...

Page 11: ...thorized to sell the product at issue within the medical animal health or dental market as the case may be in which purchaser intends to use the product 10 4 Exclusive Remedy Consequential Damages Dis...

Page 12: ...ch Email info beyes ca Web www beyes ca Printed in Canada File ENI019 Rev 3 06 01 21 Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner licen...

Reviews: