background image

4

        

Warning

• 

This module should be installed inside the dental delivery unit by autho-

rized third party technician

•  This module requirement AC24V/DC36V power, water and air from the 

delivery unit

3.1 Connection inside delivery unit

a. Board to air supply, control panel cable, motor cable and power 

cable

3. Installation requirement

Air Control

Power in AC24V/DC36

Control pannel connect

Grey
Red

Black

Yellow

Orange

Summary of Contents for Maxso E600B

Page 1: ...Electric Micromotor System Built in Instruction for Use Maxso E600B...

Page 2: ...y before operating this product for the first time please read through this manual thoroughly 1 1 Package includes instruction for use this manual user control panel control board cable o ring mountin...

Page 3: ...on and to be operated only by dentist or dental related professional Warning Install this product by authorized third party technician Follow the instruction for installation Clean and maintain this p...

Page 4: ...hnician This module requirement AC24V DC36V power water and air from the delivery unit 3 1 Connection inside delivery unit a Board to air supply control panel cable motor cable and power cable 3 Insta...

Page 5: ...5 System connect Unplug the cap b Base air and water tubing connection a Control panel b Motor 3 2 Connection outside delivery unit Air in from pedal Clear water Blue Cooling air Green spray air...

Page 6: ...am 1 Lower speed Increase speed Program 2 Keep press REV 3 seconds to open or stop c Control panel mounting Screw panel on the holder Install the plastic cap in the left or right side of the delivery...

Page 7: ...icate higher speed Step on foot paddle to start the motor During motor running light is on and speed can be adjusted 4 2 Stop Motor To stop the motor release the foot paddle Warning Do not press REV w...

Page 8: ...n motor Do not dip or wash motor Inspect o ring replace if worn or broken 5 3 Sterilization Wipe surface with ethyl alcohol Put seal blocker and seal cap on the motor Seal blocker Seal cap Put motor i...

Page 9: ...24V DC36 Motor size 22 x L71mm Motor weight 67g Noise 60dB Light source LED white light Air supply ht 1 5L min 0 2MPa Water supply 65mL min 0 2MPa Control box size L130 x W55 x H21mm Total weight 632...

Page 10: ...f invoice to the original user shall be warranted for a period of 25 months 7 3 Exclusions This limited warranty does not cover and BEYES shall not be liable for the following 1 Defects damage or othe...

Page 11: ...lth or dental market as the case may be in which purchaser intends to use the product 7 4 Exclusive Remedy Consequential Damages Disclaimer BEYES OBLIGATION UNDER THIS LIMITED WARRANTY IS THE REPAIR O...

Page 12: ...tch Email info beyes ca Web www beyes ca 0197 Printed in Canada File ENI025 Rev 2 01 03 21 Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner...