9
2.2 Auto-water supply function
2.2.1. Usage process
•
Pull out the water bottle verivally.
• Open the lid, fill the bottle with adequate purified water and then tighten the lid.
•
Clean the joint and interface of water bottle.
•
Put and place the bottle with water upside down, insert the joint into the interface vertically. (see
figure 5).
•
Press “WATER” to choose auto-water supply.
2.2.2. Notice
• Make sure the air hole and water entrance are not blocked. (see figure 8)
•
Check if the cushion inside the lid is in a good condition. If it becomes deformed or it falls off,
reinstall it or change to a new one.
•
Tighten the lid.
•
Clean the joint and interface of water bottle before each clinical operation.
• After every clinical operation with clinical liquid, change a bottle with purified water, turn the
water supply to maximum, allow the machine to work with auto-water supply for 30 seconds in
order to keep the water line and spare parts clean and durable.
•
When the water is lower than water level lower limit, add more water to keep water pipe
smooth. (see figure 9)
2.3 Endo function
2.3.1. Usage process
• Fix endochuck to handpiece by endo wrench (see figure 7)
•
Unscrew the screw cap on the endochuck.
• Put the ultrasonic file into the hole in the front of endochuck.
• Screw down the screw cap with endo wrench to tight up the ultrasonic file.
• When ultrasonic scaler turns into endo function, only the first lead light will be on and the power
will be at the 1st grade. Put the ultrasonic file into the patient’s root canal slowly, step on the
foot switch, during the treatment, turn up the power gradually according to the needs.
2.3.2 Notice
• When fixing endochuck, it must be screwed down.
•
The screw cap on the endochuck must be screwed down.