EMC DECLARATION FOR MYSCALE ANALYZER BW-SC2W AND MYSCALE ANALYZER XL BW-
SC6W
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
THE BW-SC2W AND BW-SC6W ARE INTENDED FOR USE IN THE ELECTROMAGNETIC ENVIRONMENT SPECI-
FIED BELOW. THE CUSTOMER OR THE USER OF THE BW-SC2W AND THE BW-SC6W SHOULD ENSURE THAT
THEY ARE USED IN SUCH AN ENVIRONMENT.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative hu-
midity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for pow-
er supply lines
±1 kV for input/
output lines
±2 kV for pow-
er supply lines
±1 kV for input/
output lines
Main power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differen-
tial mode
±2 kV common
mode
±1 kV differen-
tial mode
±2 kV common
mode
Main power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % Ur (>95
% dip in Ur) for
0,5 cycles
40 % Ur (60 %
dip in Ur) for 5
cycles
70 % Ur (30 %
dip in Ur) for 25
cycles
<5 % Ur (>95
% dip in Ur) for
5 sec
<5 % Ur (>95
% dip in Ur) for
0,5 cycles
40 % Ur (60 %
dip in Ur) for 5
cycles
70 % Ur (30 %
dip in Ur) for 25
cycles
<5 % Ur (>95
% dip in Ur) for
5 sec
Main power quality should be that of
a typical commercial or hospital envi-
ronment. If the user of the BW-SC2W
and the BW-SC6W requires contin-
ued operation during power mains
interruptions, it is recommended
that the BW-SC2W and the BW-SC6W
be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
EMC data for MyScale Analyzer BW-SC2W and MyScale Analyzer XL BW-SC6W
Guidance and manufacturer’s declaration - electromagnetic immunity - for
equipment and system that are not life-supporting.
Models
BW-SC2W
and
BW-
SC6W
are intended for use in the electromagnetic environment specified below. The
customer or the user of models
BW-SC2W / BW-SC6W
should assure that they are
used in such an environment.
Immunity
test
EN 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted
RF EN
61000-4-6
3 Vrms
150 kHz
to 80
MHz
N/A
Portable and mobile RF communications equipment should
be used no closer to any part of the BW-SC2W and BW-SC6W,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter. Recommended separation distance
d =1.2 √
P
d =1.2 √
P
80MHz to 800MHz
d =2.3 √
P
800MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance
level in each frequency range. b Interference may occur in the vi-
cinity of equipment marked with the following symbol:
Radiated
RF EN
61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the BW-SC2W and BW-SC6W is used exceeds the applicable RF compliance
level above, the BW-SC2W and BW-SC6W should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating
the BW-SC2W and BW-SC6W.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1]V/m.