BewellConnect - BW-TSX/BW-TS1 - User Manual - 12-2018_v2
BewellConnect - BW-TSX/BW-TS1 - User Manual - 12-2018_v2
56
57
Power
frequency
(50Hz/60Hz)
magnetic field
IEC
61000-4-8
3 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURE’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-TSX / TS1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BW-TSX / TS1 should assure that it is used in such an
environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Conducted RF
IEC 61000-4-6
Radiated RF IEC
61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.6
GHz
3 V RMS outside
the ISM band,
6 V RMS in the
ISM and amateur
radio bands 10
V/m
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the BW-TSX / TS1 including cables,
than the recommended separation
distance calculated from the equa-
tion applicable to the frequency of
the transmitter.
Recommended separation
distance
1,2
P
√
d =
1,2
P 80 MHz to 800 Mhz
√
d =
2,3
P 800 MHz to 2,5 Ghz
√
d =
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b Interference
may occur in the vicinity of
equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the BW-TSX / TS1 is used
exceeds the applicable RF compliance level above, the BW-TSX / TS1 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the BW-TSX / TS1.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE BW-TSX / TS1
The BW-TSX / TS1 is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the BW-TSX / TS1 can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the BW-TSX / TS1
as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power
of transmitter
(W)
Separation distance according to frequency of transmitter (m)
0,01
150 KHz to 80 MHz
1,2
P
√
d =
80 MHz to 800 MHz
1,2
P
√
d =
800 MHz to 2,5 GHz
2,3
P
√