15
9. What if there are problems?
Display
Cause
Solution
1. No flat, stable surface.
2. The maximum weight
capacity of 20 kg has been
exceeded.
1. Place the baby scale on a flat surface
without carpet. Re-tare the baby scale.
2. The maximum weight capacity of 20 kg
has been exceeded.
Incorrect weight
is displayed.
No flat, stable surface.
Place the baby scale on a flat surface
without carpet. Re-tare the baby scale.
Incorrect weight
is displayed.
The baby is wriggling too
much.
Try to calm the baby so that it stops
wriggling.
Incorrect weight
is displayed.
Incorrect zero point of scale.
Switch the baby scale off and on again.
Wait until the baby scale displays "
0.000
kg"
and repeat the measurement.
The batteries in the scale are
empty.
Replace the batteries in the baby scale.
10. Disposal
For environmental reasons, do not dispose of the device in the household waste at the end
of its useful life.
Dispose of the device at a suitable local collection or recycling point in your country.
Dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and
Electronic Equipment). If you have any questions, please contact the local autho rities
responsible for waste disposal.
Battery disposal
Empty, completely flat batteries must be disposed of through specially designated collec-
tion boxes, recycling points or electronics retailers. You are legally required to dispose of
the batteries.
The codes below are printed on batteries containing harmful substances:
Pb = Battery contains lead
Cd = Battery contains cadmium
Hg = Battery contains mercury
11. Technical specifications
Model:
BY 80
Dimensions:
56cm x 33cm x 5.4cm (W x D x H)
Weight:
approx 1.7kg
Measurement range:
0.05kg to 20kg
Scale interval d:
The display can be read in increments of 0.005kg
Repetition accuracy:
The measuring tolerance for repeated measurements is +/- 1% (several mea-
surements in direct succession on the same scale with position of scale and
person the same as far as possible)
Absolute precision:
In comparison to a calibrated weight, the measured value is +/-0.2kg.