12. Technical specifications
Model no.
BM 26
Measurement
method
Oscillometric, non-invasive blood pressure
measurement on the upper arm
Measurement
range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
pulse 30 – 180 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5% of the value shown
Measurement
uncertainty
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /
diastolic 8 mmHg
Memory
4 x 30 memory spaces
Dimensions
L 155 mm x W 110 mm x H 70 mm
Weight
Approximately 395 g
(without batteries, with cuff)
Cuff size
22 to 35 cm
Permissible
operating
conditions
+10°C to +40°C, 15-90% relative humidity
(non-condensing)
Permissible
storage conditions
- 25°C to + 70°C, ≤ 93% relative humidity,
700 – 1060 hPa ambient pressure
Power supply
4 x 1.5 V
AA batteries (alkaline
type LR6)
Battery life
For approx. 180 measurements, depending
on levels of blood pressure and inflation
pressure
Classification
Internal supply, IPX0, no AP or APG, con-
tinuous operation, application part type BF
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This device conforms with the European standard
EN 60601-1-2 (in accordance with CISPR 11, IEC 61000-3-2,
IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3,
IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6,
IEC 61000-4-8, IEC 61000-4-11) and is subject to particu-
lar precautions with regard to electromagnetic compatibili-
ty. Please note that portable and mobile HF communication
systems may interfere with this device.
• This device complies with the EU Medical Devices Directive
93/42/EEC, the German Medical Devices Act (Medizinpro-
duktgesetz) and the standards EN 1060-3 (Non-invasive
sphygmomanometers – Part 3: Supplementary requirements
for electro-mechanical blood pressure measuring systems)
and IEC 80601-2-30 (Medical electrical equipment – Part 2 –
30: Particular requirements for the basic safety and essential
performance of automated non-invasive sphygmomanom-
eters).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful
life. If the device is used for commercial medical purposes, it
Summary of Contents for BM 26
Page 82: ...82 Beurer 1 83 2 83 3 84 4 84 5 86 6 87 7 88 8 91 9 92 10 92 11 92 12 93 13 94...
Page 84: ...84 SN CE 93 42 EEC 3 BM 26 22 35 4 30 5 1...
Page 85: ...85 3...
Page 86: ...86 5 1 2 3 4 5 6 M 7 SET 8 9 10 SYS mmHg DIA mmHg PULSE min SET M 4 9 10 1 2 3 6 7 8 5...
Page 88: ...88 SET 7 1 2 2 3 3 4 5 OK 22 35 30 42 162 973...
Page 89: ...89 5 1 SET 30 M 30 Err...
Page 90: ...90 3 1 5 3 180 110...