130
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The BM 55 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 55 should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage dips, short inter-
ruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the BM 55
requires continued operation during power mains inter-
ruptions, it is recommended that the BM 55 be powered
from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for BM 55
Page 97: ...97...
Page 98: ...98 WEEE Waste Electrical...
Page 102: ...102 071 60 5 2 3...
Page 103: ...103 22 36 35 44 163 387 5 START STOP 3 START STOP...
Page 104: ...104 M1 M2 START STOP 5 6...
Page 106: ...106 5...
Page 108: ...108 M1 M2 03 M1 M2 START STOP START STOP M1 M2 5 8 USB PC beurer HealthManager 1 2 30 9 1 2...
Page 109: ...109 10 _ Lo 300 160 Lo IEC 60601 1 8 8...
Page 111: ...111 IEC80601 2 30 2 30 12 FW 7575M EU 6 06 100 240 50 60 6 600 Beurer 2...
Page 112: ...112 13 36 5 218 89077 OOO 109451 62 2 3 109451 62 2 495 658 54 90 bts service ctdz ru...