Electromagnetic Compatibility
Information
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The BC 28 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BC 28 should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 2
The BC 28 must emit electromagnetic energy in order to perform its inten-
ded function. Nearby electronic equipment may be a
ff
ected.
RF emissions CISPR 11
Class B
The BC 28 is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Not applicable
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The BC 28 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 28 should
assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteri-
stic of a typical location in a typical commercial or hospital envi-
ronment.