![Beta 1760/2TM Manual Download Page 3](http://html.mh-extra.com/html/beta/1760-2tm/1760-2tm_manual_2742498003.webp)
ENEN
3
ELECTRICAL SPECIFICATIONS
Temperature Scale:
Celsius (
°
C) or Fahrenheit (T) user-selectable
Measurement Range:
-50
°
C to 1300
°
C -58
°
F to 2000
°
F
Resolution:
1
°
C or 1
°
F, 0.1
°
C or O.1
°
F
Accuracy:
Accuracy is specified for operating temperatures over the range of
18
°
C to 28
°
C (64
°
F to 82
°
F), for 1 year, not including thermocouple error.
±
2
°
C
-50
°
C to 0
°
C
±
4
°
F
-58
°
F to 32
°
F
±
(0.5%rdg+1
°
C)
0
°
C to 1000
°
C
±
(0.8% rdg+1
°
C)
1000
°
C to 1300
°
C
±
(0.5% rdg+2
°
F)
32
°
F to 2000
°
F
Temperature Coefficient:
0.1 times the applicable accuracy specification per
°
C
from 0
°
C to 18
°
C and 28
°
C to 50
°
C
(32
°
F to 64
°
F and 82
°
F to 122
°
F).
Input protection:
60V dc or 24V rms ac maximum input voltage on any combination
of input pins.
Reading Rate:
2.5 times per second.
Input Connector:
Accepts standard miniature thermocouple connectors (flat blades
spaced 7.9mm, center to centier).
Environmental
Ambient Operating Range:
0
°
C to 50
°
C (32
°
F to122
°
F)
Storage Temperature:
-20
°
C to 60
°
C (-4
°
F to140
°
F)
Relative Humidity:
0% to 80% (0
°
C to 35
°
C) (32
°
F tc 95
°
F)
0% to 70% (35
°
C to50
°
C) (95T to122
°
F)