TECHNICAL SPECIFICATION :-
4
INSTRUCTIONS FOR USE
Display:
4 digit, 7 segment LED.
Display update rate:
2 readings per second (2Hz).
Zero suppression:
± 2 least significant digits
Auto reset hold time:
2 seconds.
Operating temperature range:
Storage temperature range:
+5°C to +40°C.
-20°C to +70°C.
Maximum operating humidity:
85% Relative humidity @30°C.
AC power adapter:
230 Volts AC at 50 Hz input (UK / Europe).
240 Volts AC at 50 Hz input (Australia).
120 Volts AC at 60 Hz input (USA / Canada).
90 to 264 Volts AC at 50-60 Hz input. (World).
6V, 300 mA DC output (centre positive)
Power consumption:
1.8 W - maximum.
Weight:
2 kg shipping weight.
Dimensions (in mm):
175 (L) x 63.5 (W) x 63.5(H)
Case materials / finish:
Powder coated aluminium housing. Stainless steel tran-
sducer shaft..
Environment:
Indoor use within a light industrial environment.
Electromagnetic Compatibility (EMC) Directive:
In conformance with EN 61326-1 : 2006.
Low voltage directive:
In conformance with EN 61010-1 : 2001.
To environmental conditions Pollution Degree 2
& Installation Category (Over voltage Category) II.
Due to continuous improvement all specifications are subject to change without prior notice.
CALIBRATION :-
Your instrument has been supplied with a declaration of conformity. To maintain the specified
accuracy it is recommended that the instrument is recalibrated at least once per year. Recalibration
should be carried out by the Supplier or by the Supplier approved agent, where all the
facilities to ensure the instrument is functioning at maximum accuracy are available.
Do not remove front panel or case as there are no calibration settings inside.
REPAIR :-
Repair should be carried out by the Supplier or by the Supplier approved agent, where all the
facilities to ensure the instrument is functioning at maximum accuracy are available.
There are no parts for user repair inside the case.
CLEANING:-
Do not use abrasives or solvent based cleaners.
WARNING:-
If the instrument is used in a manner not specified by the Manufacturer, the protection provided by
the equipment may be impaired.