1. Aim and use

1.1 Purpose (Section of the user instruction)

BestLift Luna is a ceiling hoist designed to transfer patient in institutions such as nursing homes,

hospitals, care centres and in home care. The BestLift Luna ceiling hoist can be used as a ceiling 

mounted lift or as a mobile lift in fixed or movable track systems.

The use of the BestLift Luna requires the following:

                  

Personnel have received qualified instruction by Bestcare in using the BestLift Luna.

                  BestLift Luna is used in suitable track systems that have been approved by Bestcare.

                  System assembly is carried out by installers authorized by Bestcare.

                  BestLift Luna can only be fitted with spreader bars supplied by Bestcare. Other spreader

                  bars can only be used if specifically approved and authorized in writing by Bestcare.

 

    The BestLift Luna system has been developed for use, with Bestcare lifting slings and 

 

    other sling products authorized by Bestcare. See Bestcare’s sling leaflet for specific 

 

    information about authorised models.

 

    BestLift Luna can be mounted direct on tracks or mounted in strap. If the lift is required in

 

    a different mode to that delivered then your dealer or Bestcare’s service team must be 

 

    contacted.

1.2 Safety instructions

This description should be read carefully before BestLift Luna is taken into service.

 

    

All parts must be mounted correctly according to this guide.

 

    The BestLift Luna must be checked at least once a year according to EN/ISO 10535.

 

    The BestLift Luna lifting capacity is 275KG/606LB depending on the model and

 

    must never be exceeded. When using BestLift Luna in connection with other componets

 

    e.g. slings or transportable track systems it is always the component with the LOWEST

 

    weight capacity that decides the maximum lifting capacity.

1.3 Daily check (Section of the user instruction)

The carer/helper is reponsible in ensuring that a visual check is carried out BEFORE BestLift Luna

and asscessories are used, including the following:

 

    

 

   

 Lifting slings are checked for wear and tear or any damage especially as regards lifting

 

    straps. The slings must not be used if defects are detected.

 

    The lifting strap on the BestLift Luna is checked for wear and tear or damage. If there is

 

    any sign of a defect the BestLift Luna must be taken out of use.

 

    If any abnormal sound is heard from the BestLift Luna, the hoist must not be used before

 

    personnel authorized by Bestcare have carried out a safety check on the unit.

 

    Check tat the lifting strap moves into the lifting unit without touching the edges of the hole.

 

    If in any doubt concerning the BestLift Luna or lifting slings contact Bestcare for advice and

 

    guidance.

Summary of Contents for BestLift Luna

Page 1: ...BestLift Luna Ceiling hoist Technical service Date 02 26 2015 Version 1 ...

Page 2: ...1 ...

Page 3: ...2 1 Aim and use 1 1 Purpose 3 1 2 Safety instructions 3 1 3 Daily check 3 1 4 Terminology 4 2 A nnual check and ser vice 2 1 Annual check 5 2 2 Servi ce check 6 ...

Page 4: ...ions This description should be read carefully before BestLift Luna is taken into service All parts must be mounted correctly according to this guide The BestLift Luna must be checked at least once a year according to EN ISO 10535 The BestLift Luna lifting capacity is 275KG 606LB depending on the model and must never be exceeded When using BestLift Luna in connection with other componets e g sling...

Page 5: ...fixing for spreader bar Carry handle Bayonet on spreader bar Hook for sling lifting strap Track system Function buttons and connection of hand control Strap Fig 1 4A Carry handle Hook for removable spreader bar Track system Function buttons and connection of hand control Strap Hook for sling lifting strap ...

Page 6: ...er bar and lifting bolt check for wear and tear or deformation Electric parts check emergency stop emergency lowering all plugs and functions are tested Test loading the lift must be tested for maximum load and critical points checked Parts for lifting patients Slings and straps all stitching and sewing points are checked look for any holes or wear and tear Important Remember to fill out the servic...

Page 7: ...n position Lifting hook must move and close easily and must also audibly click when locking Check that the holding pin is in place a Check for any abnormal sounds up down with and without load The hoist is loaded to a test load of 275 kg 606LB ensure that the track is approved to this weight Lift the load approx 5cm and check the hoist and track for any faults or deformations Move the hoist over t...

Page 8: ...7 93 42 EØF CLASS I EN ISO 10535 EN 60601 1 2 Phone 678 679 6690 Fax 678 679 6697 www bestcarellc com Manufacturer Bestcare LLC 2920 Pacific Drive Norcross GA 30024 USA ...