-2-
2.
Specification
RCD Testing
Range
10mA±4mA Trip,14mA↑ over,150mS±10%
30mA±12mA Trip,42mA↑ over,150mS±10%
150mA±60mA Trip,210mA↑ over,30mS±10%
Insulation Testing
Range
1MΩ / 9.9MΩ / 99MΩ
Accuracy
± 1%
Max. Testing
1.2kV
Continuity Testing
Range
0.5Ω / 2Ω / 10Ω
Accuracy
± 0.05Ω / ± 0.1Ω / ± 0.2 Ω
Max. Current
300mA
LOOP Testing
Range
Supply Loop / Supply Loop + 1Ω
Accuracy
0.45Ω ± 0.05Ω / 1.45Ω ± 0.05Ω