5. Electrical Fault Finding
(Continued)
5.2
Coolant Pump
If the voltage is correct at the coolant pump a new pump is required.
5.3
Traverse Motor
If the voltage is correct at the traverse motor terminals, try fitting a new capacitor before
assuming that the motor itself is at fault.
If the traverse jumps at the change in direction, check microswitch and contactor operation
but most likely is a capacitor failure (often resulting from a low voltage supply).
5.4
Grinding Motor
If the voltage is correct at motor terminals check the current/amps at terminals T2 & T3 (on
thermal overload contactor, then at “MAIN” plug on Control PCB). If the current exceeds the
full load current indicated on the motor plate, a new motor is needed. If below full load current
the overload could be set too low (see Service Bulletin #4).
5.5
Magnets
Check for 12V DC at terminals 2 and 3 at the 4 pole socket at the rear of the drawer (closest
to the largest 16 pole connector. If no volts are present, check output at “MAGNET” plug on
main Control PCB - then at transformer (
Warning
: where voltage will be AC).
17
ANGLEMASTER • AM3000
© Bernhard and Company Limited
Summary of Contents for AM3000
Page 20: ...7 Parts List Continued BASE FRAME 20 ANGLEMASTER AM3000 Bernhard and Company Limited...
Page 22: ...7 Parts List Continued MAIN FRAME 22 ANGLEMASTER AM3000 Bernhard and Company Limited...
Page 24: ...7 Parts List Continued DUST TRAY GUARDS 24 ANGLEMASTER AM3000 Bernhard and Company Limited...
Page 28: ...7 Parts List Continued 28 ANGLEMASTER AM3000 Bernhard and Company Limited...
Page 30: ...7 Parts List Continued MAIN MOTOR AND SLIDE 30 ANGLEMASTER AM3000 Bernhard and Company Limited...
Page 36: ...8 Wiring Diagrams Continued SHEET 2 36 ANGLEMASTER AM3000 Bernhard and Company Limited...
Page 37: ...8 Wiring Diagrams Continued SHEET 3 37 ANGLEMASTER AM3000 Bernhard and Company Limited...