10.1. Alarm display and confirmation
options
Option 1: Alarm indication “all alarms”
With this option the alarms will be visible on the display with an alarm symbol
for 30 days.
By pressing the READ button, the warning symbol
will be disabled for the corresponding alarms. The
alarm symbol
cannot be canceled nor reset.
Note:
• In this mode only one upper and one lower alarm will be triggered per day.
• The alarm symbol
will be present on the display for 30 days.
• The warning symbol
can be deactivated by confirming all existing alarms in the readout mode.
• The alarm buzzer stops when the alarm is confirmed within the set alarm limits. Otherwise the
buzzer pauses for approx. 1 hour and starts again for up to 168 hours (7 days).
Option 2: Alarm indication “unconfirmed alarms”
The alarms are shown with the alarm symbol
until all alarms (in the 30-day history) have been
Berlinger & Co. AG
User Manual Fridge-tag Ultra Low - 1
Page 31 of 57