
Chapter 15 Troubleshooting and correcting faults
15 Troubleshooting and correcting faults
HOTLINE
Notify Berlin Heart! 866.249.0128
Problem
Cause of problem / action to be taken
Deposits in the
Initial deposits: check anticoagulation status and adjust
pump
therapy if necessary.
If floating deposits are detected (may cause
thromboembolic complication): replace the pump, see
section 15.1: Replacing the blood pump(s), page 195.
Visible blood pump
faults
Replace the pump, see section 15.1: Replacing the blood
pump(s), page 195.
Pump membrane
Possible causes:
remains in the
diastolic or systolic
position despite
vibration /
• kinking of the cannula
• clotting of the pump
• partial malfunction of the Ikus
movement of the
What to do?
pump indicating
Check for external forces on the cannula and whether it
that the Ikus is
may be necessary to manipulate the cannula, see
attempting to
section 14.4: Please check left / right pump and driving
provide diastolic or
systolic pressure
Check for clots in the pump or cannula that may be
obstructing flow and replace the pump if necessary, see
section 15.1: Replacing the blood pump(s), page 195.
Initiate hand-pumping to try to eject the pump, see
section 15.5: Driving blood pump(s) with the manual
pump, page 204.
Switch the patient to the back-up Ikus driving unit, see
section 15.4: Connecting the patient to a replacement
Ikus, page 202.
Pump membrane
Additional possible causes:
remains in one
position despite the
above
• High vascular resistance
• Defective blood pump
manipulations
There may be air leaking into the space between the first
and second layer of the triple-layer pump membrane. This
accumulated air may gradually create a “pillowing” effect
between the membranes.
The top (visible) membrane layer will appear to be
continuously in diastole while the bottom two membrane
layers are in fact continuously in systole.
Tab. 15-1
Possible problems
1000721x09 Revision 8
191
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...