Chapter 14 Error Messages and corrective measures
INSTRUCTION
1.
Check the status of the patient.
2.
Evaluate the malfunction scenario:
•
no alarm, but the Ikus continues to function:
no communication
between the control computers and the laptop
•
optical and acoustic alarm, and the Ikus is working in emergency
mode:
both control computers are defective
•
The Ikus is halted; acoustic alarm only:
Defective power supply
3.
Provide the patient immediately with a replacement Ikus and switch off the
malfunctioning Ikus. (see section 15.4: Connecting the patient to a replace
ment Ikus, page 202).
If no replacement Ikus is available and the Ikus is running: Check the filling
and emptying of the blood pump(s), and monitor the function of the Ikus. Do
not operate the Ikus without supervision!
If no replacement Ikus is available and the defective Ikus is halted: Support the
patient with the manual pump. (see section 15.5: Driving blood pump(s) with
the manual pump, page 204).
HOTLINE
Notify Berlin Heart! 866.249.0128
14.22 Left/right flow sensor defective. Notify Service!
•
Left flow sensor fault. Contact customer service.
•
Right flow sensor fault. Contact customer service.
The corresponding flow sensor is defective. Although the Ikus continues to run, the
excessively low flow would not be detected - possibly due to a kink in a cannula or the
driving tube.
WARNING
Do not operate the Ikus without supervision! Otherwise an insufficient
support of the patient might not be detected.
NOTICE
In biventricular mode there is a flow alarm for one of the blood pumps.
The user stops the pump operation of the other blood pump and
restarts it. This affect that during the restart the flow alarm will be
deleted, wether it’s reason continue to exist or not.
IMPORTANT: If the defective Ikus is not replaced, this message appears repeatedly
at 10-minute intervals.
INSTRUCTION
1.
Check the status of the patient.
2.
Check the filling and emptying of the blood pump(s), and monitor the function
of the Ikus. Do not operate the Ikus without supervision!
1000721x09 Revision 8
185
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...