Chapter 14 Error Messages and corrective measures
3.
Reset all of the parameter values. In addition, log out of the monitor program
and then log back into it. Navigate the cursor with
<
←
>
/
<→>
to the desired
field, adjust the value with
<↓>
/
<↑>
/
<Bild
↓
>
,
<Bild
↑
>
, then confirm with
<Enter>
. The system works with the new value..
If the message appears again:
HOTLINE
Notify Berlin Heart! 866.249.0128
14.11.4... reports faulty test
Backup computer reports faulty test
The main computer cannot end the test phase to check the system after the error
message...(See section 14.1: Pressure error / time error in system 1 (or in system 2
or 3), page 173.)
•
Pressure fault in system 1 (left), please wait ...
•
Pressure fault in system 2 (right), please wait ...
•
Pressure fault in system 3 (backup), please wait ...
•
Time error system 1 (left), please wait ...
•
Time error system 2 (right), please wait ...
•
Time error system 3 (backup), please wait ...
IMPORTANT: The blood pump being driven by the tested system stops for approx. 10
seconds during the test phase.
INSTRUCTION
1.
Check the status of the patient.
2.
If the Ikus switches to backup operating mode or this message repeatedly
appears, inform Service immediately.
3.
Provide the patient immediately with a replacement Ikus. See section 15.4:
Connecting the patient to a replacement Ikus, page 202.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.12 Measurement discrepancy in main computer (backup
computer)
•
Measurement discrepancy in main computer!
•
Measurement discrepancy in backup computer!
WARNING
Between the error message
Measurement discrepancy in main
computer!
and the 2nd (relevant) error message there can be a delay
of several seconds. At any case, wait for both error messages.
Only the active processor has detected an error. This message only appears in
conjunction with an additional, relevant error message. Immediately take all of the
necessary measures for this second message.
If the message appears only once at the backup computer,
it is only an information message.
1000721x09 Revision 8
180
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...