Chapter 14 Error Messages and corrective measures
14.1 Pressure error / time error in system 1 (or in system 2 or 3)
•
Pressure fault in system 1 (left), please wait ...
•
Pressure fault in system 2 (right), please wait ...
•
Pressure fault in system 3 (backup), please wait ...
•
Time error system 1 (left), please wait ...
•
Time error system 2 (right), please wait ...
•
Time error system 3 (backup), please wait ...
The air volume pumped per cycle by the Ikus or the pressure required to pump it has
changed. The Ikus checks whether there is an internal system fault (e. g. a
compressor has broken down) or whether there is an external error (e. g. driving tube
not connected or driving tube leak).
WARNING
The blood pump of the corresponding system is stopped for the
duration of this test (for approx. 10 seconds).
If the Ikus has detected a pneumatic system fault…
the Ikus switches to backup operation and the message Backup operation left/right.
Contact customer service. is shown (see section 14.5: Backup operation left/right,
page 176), together with a request to inform the service department.
HOTLINE
Notify Berlin Heart! 866.249.0128
If the Ikus has detected an external fault …
it displays the message Please connect driving tube! (see section 14.3: Please
connect driving tube, page 174). Provide the patient immediately with a replacement
Ikus.
If the main computer and the backup computer arrive at different results in the
test phase …
the backup computer generates the message Backup computer reports faulty test
(see section 14.11.4: ... reports faulty test, page 180).
Follow the instructions exactly as described in each section.
14.2 Throttle valve error in system 1 (or system 2 or 3)
•
Throttle fault in system 1 (left), please wait ...
•
Throttle fault in system 2 (right), please wait ...
•
Throttle fault in system 3 (backup), please wait ..
.
HOTLINE
Notify Berlin Heart! 866.249.0128
In an active system, the Ikus has detected a throttle valve error (throttle: internal
module for pneumatics test) and switches to backup operation (see section 14.5:
Backup operation left/right, page 176). Provide the patient immediately with a
replacement Ikus. A throttle is an internal assembly for the pneumatic test.
1000721x09 Revision 8
173
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...