28
1015002x00
Safety Information
en
WARNING
WARNING
Only connect electrical components to fixed mains power outlets. Do
not use extension cables. Take into account the protection class of the
power outlet and components!
Do not touch the electrical contacts on a component and the patient at
the same time. Doing so could cause contact voltage to reach an intra-
cardial area. which could harm the patient.
4.7
Power Supply
When switching from mains to battery operation, check the charge level
of the batteries. If a longer period of non-mains operation is anticipated,
check the charge levels of the replacement batteries as well. Do this by
pressing the
<charge level button>
on the battery. If a power supply
is not available, the driving unit will stop.
When using the system in mobile operation, always ensure that an al-
ternative source of power can be reached within the battery operating
time. Keep in mind that batteries can also fail. Alternative sources of
power could include: Replacement batteries, extra power supply unit
with means of connecting to mains power.
4.8
Operation
Ensure that driving unit audible alarms can be reliably heard by medical
personnel / caregivers at all times. Other methods of monitoring the
driving unit, such as the external alarm (nurse call), must only be used
as supplementary tools.
The values and data displayed assist in the monitoring and configura-
tion of the EXCOR Active system. Other therapeutic measures must al-
ways be reinforced through appropriate diagnostic procedures.
Therapeutic measures must not be derived solely on the basis of the
values and data shown on EXCOR Active.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...