124
1015002x00
Replacing the Driving System
en
9.2.3
Switching to EXCOR Active
3.
Configure the driving unit, see section 8.5: Configuring the System, page 98.
Transfer the following parameters from Ikus to EXCOR Active (for each sup-
ported side in BVAD mode):
•
Operating mode
•
Cannula sizes
•
Size of the blood pump
•
Rate
•
Synchronization
type (in BVAD operating mode)
•
Initially, leave the parameter
Duration
systole set to the standard value of
50%.
4.
Start operation of the blood pump(s) on the driving unit.
5.
Position the driving unit on the Ikus so as to allow fast and secure reconnec-
tion of the driving tube.
1.
Lay the patient on his or her back. Perform the following steps (for each sup-
ported side in BVAD mode):
2.
Connect the flow sensor to the cannula.
3.
Connect the flow sensor to EXCOR Active.
4.
Remove the seal plugs from the EXCOR Active driving unit.
5.
Switch the driving tube connection from Ikus to EXCOR Active.
6.
Check filling and emptying of the blood pump and adjust as needed.
7.
Set the
Flow alarm threshold
.
8.
Attach the cable wrap.
9.
Allow Ikus to remain active for a while, in parallel with the EXCOR Active driv-
ing unit. Switching back is easy to do if necessary. Connect the tank unit so
that Ikus does not run with the pneumatic connections open.
10.
Adjust the parameters for mobilization purposes, see section 8.8.3: Mobilizing
Patients, page 111.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...