Configuring and Operating the Driving Unit
en
8.8.3
Mobilizing Patients
Because of the changing hemodynamic conditions when lying down, sitting, standing,
and with normal movement, perform mobilization gradually.
Checking support while sitting
Checking support while standing
Operating mode
Operating modes can only be
changed after the driving unit is
reset.
Cannula sizes
Cannula sizes can be only
changed after the driving unit is
reset.
Parameter
Auto operation
Blood pump(s) stopped
Tab. 8-4
Changing parameters
1.
Initial state:
Driving unit and panel PC connected, both devices are on.
2.
Log in -
E
XPERT
OR
Medical.
3.
Select
B
LOOD
PUMP
settings or
D
RIVING
UNIT
settings.
•
Duration systole
•
Rate
•
Flow alarm threshold
•
Filling / Emptying
•
Synchronization
•
Patient weight
•
Language for driving unit
•
Child protection
•
Time of self-test
4.
Choose and set parameters that are to be changed.
5.
When changing blood pump parameters, pay attention to the correct filling
and emptying of the blood pump. Observe patients.
1.
Ask the patient to sit on the edge of the bed for 5 to 10 min and check the fill-
ing behavior of the blood pumps.
2.
If filling / emptying worsens, adjust the parameters.
1.
Ask the patient to get up out of bed for 5 min.
2.
If filling / emptying worsens, adjust the parameters.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...