Configuring and Operating the Driving Unit
en
WARNING
8.7.2
Setting a Flow Alarm Threshold
Set the flow alarm threshold for each patient individually; check it and
adjust it as needed. If the flow alarm threshold is set too low, it can re-
sult in insufficient patient support not triggering alarms. Do not set the
flow alarm threshold to a value that will make the alarm unusable.
1.
The flow alarm needs to signal insufficient support reliably, while avoiding
unnecessary alarms for minor flow fluctuations.
The flow alarm must be set immediately after connection of the flow sensor.
Each time after the rate is adjusted, check the flow alarm threshold and adjust
it if necessary.
2.
Call up the
B
LOOD
PUMPS
/P
ARAMETERS
view. Set a separate
Flow alarm
threshold
value for each flow sensor.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...