FHIT TEC_FHIT_M Rev.1 11/2016
33
IEC 61000-4-11
40% U
T
(60% hole in U
T
for 5
cycle)
70% U
T
(30% hole U
T
for 20
cycle)
<5% U
T
(>95% hole U
T
for 5S)
70% U
T
20 cycle (500mS)
<5% U
T
5S
interruption, we recomend
device fueling with a
uninterruptible power supply
or with battery.
Frequency magnetic
field (50/60 Hz)
IEC 61000-4-8
3A/m
3A/m
Frequency magnetic field
must have the same
characteristic as the tipical
location or commercial or
hospital environment.
NOTE: U
T
is the line voltage in c.a. before the test level application.
Guide and manufacturer declaration
– electromagnetic immunity
The device has to operate in the electromagnetic einvironment below specified. Devie customer or user
must guarantee that the device is used in this environment.
Immunity test
Test level IEC 60601
Conformity
level
Electromagnetic einvironment -
guide
Pipe RF
IEC 61000-4-6
Radiate RF
IEC 61000-4-3
3 Veff
From 150 kHz to 80
MHz
3 V/m
From 80 MHz to 2,5
GHz
RF comunication device, portable
and fixed musten’t used in the
device part proximity, also cable,
from the recomended separation
distance recomended, calcolated
with the equation applicable to the
trasmitter frequency.
Separation distance recomended:
d=1,17
√
d=1,17
√
from 80 MHz to 800
MHz
d=2,34
√
from 800 MHz to 2,5
GHz
P
is the max nominal exit potential
of trasmitter in Watt (W) according
to the trasmitter manufacturer and
d
is the separation distance
recomended, in meter (m).
RF fixed trasmitter field intensity,
definied with an electromagnetic in
loco survey, must be lower than the
comformity level for each
frequency
b
interval.
It could be test the inerference in
proximity of device with this symbol:
NOTE1: 80 MHz and 800 MHz, is applied the separation distance for the highr frequency interval