SECOND PART- fHIT EXAM
PREPARATION
4.1.1.1.1.3.
Warning !: Do not practice impulses to the head of patients with neck injuries or
patients that the doctor has advised to limit or avoid neck movements
4.1.1.1.1.4.
Connect the sensor with a 3 meters USB cable.
Switch on the sensor using the button next to the entrance of the USB cable on the sensor; to turn it on you
may use a ballpoint pen
Position the patient at 150cm from the computer screen or the same distance used for the examination of
visual acuity.
Protect the area of head where the elastic band with the sensor is positioned using a sterile gauze.
Tie the sensor on the patient's head, making sure that the three LEDs are on the top.
The lights that flash on the sensor refer to its movement. The lower blue light indicates that the sensor is
turned on. The red light in the center is used to indicate the proper functioning of the sensor
Bring the cables behind the patient to avoid obstructing the examination or generate false results.
The physician stands behind the patient holding his hands between the jaw and the patient's cheeks, taking
care to avoid touching the sensor band.
Pressing the CONTINUE button that will appear on the following screen that verifies the correct USB
connection.