Case Expansion Monitor Maintenance
6 I Monitor Options [Cont]
Table 1 [continued]
84418-01
•
Option
Setting
Jumper
Install R e m o v e
Upscale Channel A T o w a r d Transducer *W12C
Away from Transducer
W
1
2
C
Upscale Channel B
Toward Probe * W 1 2 B
Away from Probe
W
1
2
B
Note : Both channels must be configured in the same direction. An error code 22 will occur if both
channels are not configured the same. See section 15.
Recorder Output Options
To set the recorder output, remove all jumpers from headers W1A-E, W2A-E, W3A-D, W5A-C, W6A-
C, W8A-D, W9A-D, W14G AND W14H. Install jumpers specified in Table 2.
Table 2. Recorder Output Options
Recorder Output
+
4
to +20 mA *
+1 to + 5 Vdc
0
to -10 Vdc
Jumper Settings
W1B,C,E
W2B,C,E
W3A D
'
W5A
W6A
W8B,C
W9B,C
* Options as shipped from factory
W1A,D
W2A,D
W3A,D
W5C
W6C
W8D
W90
W14H
W1A,D
W2A,D
W3B,C
W5B
W6B
W8A
W9A
W14G
Option
Setting
Jumper
Install R e m o v e
Recorder Mode
Differential * W 1 2 E
Direct
W12E
10
•
•
Summary of Contents for 3300/48
Page 5: ...84417 01 C a s e Expansion Monitor Operation Blank Page i v...
Page 7: ...84417 01 Case Expansion Monitor Operation Blank Page v i...
Page 25: ......
Page 29: ...Case Expansion Monitor Maintenance Blank Page i v...
Page 31: ...Case Expansion Monitor Maintenance 8 4 4 1 8 0 1 Blank Page v i...
Page 72: ...Case E x p a n s i o n M o n i t o r S c h e m a t i c Sheet 5 o f 11...