Syncing with the light source’s color temperature
You are recommended to adjust the display’s color temperature before using an endoscope.
1. Make sure the endoscope and its light source have been properly connected with the dis-
play, and then go to
System
>
Endo light
to enable the light sensor on the bottom of the
front bezel. Check if there are any obstacles in front of the sensor that keep it from detect-
ing the light properly.
2. Turn on the light source, and aim the endoscope at the sensor with 3 ~ 5 cm of distance.
Follow the on-screen instructions to make adjustment if necessary.
3. After the measurement is done, press
MENU
to exit.
Managing multiple windows (
PIP/PBP
)
You are provided with different PIP/PBP modes and several ways to manage the screens.
Selecting input sources
1. Make sure the signal cables have been connected properly.
2. Press the
ENTER
key to bring up a list of all input sources. Follow the control tips
displayed on the bottom of the list to select the source for the main and the sub screens
separately. Depending on the selected main source, some of the sources are available for
the sub screen.
The sensor works within certain color temperature range only and may fail to complete the measurement if it is
not within this range.