34
35
Chapter 9 Specifications
System Accuracy : ± 20% at glucose level > 75 mg/dL
Test Sample : Capillary whole blood
Measuring Time : Less than 10 seconds
Measuring Range : 20 - 600 mg/dL (1.1 - 33.3 mmol/L)
Sample Volume : About 1 - 1.5 µL
Strip Storage Condition :
10 - 30˚C
Operation Temperature :
10 - 40˚C
Relative Humidity : Less than 95%
Memory : 180 test results
Battery Type : One 3V (CR2032) lithium battery
Battery Life : Approximately 1,000 tests
Dimensions : 61*77*19 mm
Weight : About 47g (batteries not included)
Message
Cause
Solution
Hi
1.Blood glucose level is
higher than 600 mg/dL.
2 .Test strip has problem.
1. Review the instructions and try
again with a new test strip.
2. Check the test strip vial to make sure
that it is not marked beyond the
expiration date.
Lo
1.Blood glucose level is
lower than 20 mg/dL.
2.Test strip has problem.
1. Review the instructions and try
again with a new test strip.
2. Check the test strip vial to make sure
that it is not marked beyond the
expiration date.
03-RN01-UM-01G BK使用手冊(英文)(文字檔)-234-35 34-35
2009/8/21 下午 04:51:21
Summary of Contents for GLU
Page 1: ......