e.g.: line1/line 2 ready for operation, Q1 on / Q2 on. The values can be assigned via the 
menu of the control device (see instruction leaflet PRC487).

•  Switching back interlocking: On recovery of the voltage, switching back to the preferred 

supply line can be disabled or enabled via an external contact. Please observe the requi-
rements of DIN VDE 0100-710 (VDE 0100 Part 710):2002-11 Abs. 710.537.6.1.f)

3.3 

Instruction for connection

3.3.1  Alarm indicator and test combination

Ex works, the following pairs of terminals are provided for the connection of BMS bus devices:  
 

X3:29 and X3:30 (A/B) 

 

X3:31 and X3:32 (A/B terminals are terminated with 120 Ω at the factory)

Alarm indicator and test combinations, panels and other Bender-BMS-bus devices can be connec-
ted.

1. A BMS device or an existing BMS bus with several devices is connected to terminals 

X3:29 and X3:30: 
The last device at the other end of the bus must be terminated with 120 Ω. Terminals 
X3:31 and X3:32 remain terminated.

2. An existing BMS bus that has already been terminated on both ends, is disconnected; 

one open branch is connected to terminals X3:29 and X3:30, while the other one is con-
nected to terminals X3:31 and X3:32:  
The 120 Ω resistor, provided ex works, must be removed and the open bus branches 
must be connected to the specified terminals.

Please read the information on cable routing in the „BMS bus“ instruction leaflet.

The AN450 power supply unit can supply power to a maximum of 2 MK2430 alarm indicator and 
test combinations via the terminals X3:33 and X3:34. In this respect, please refer to the documen-
tation for the relevant devices. The AN450 is not suitable for supplying power to TM... operator 
panels.

3.3.2  TMX-HA

Connect the alarm indicator and operator panel TMX-HA to the terminals X3:39 ... X3:70 of the 
changeover module UMC . 710D4-..-HA. The panel is supplied via the power supply unit AN450 
via the terminals 33 and 34.

3.3.3  GLT/ZLT

If messages from the UMC710D4-.. changeover module are to be transmitted to a Building 
Control System or to a Central Process Control System, you have the following options:

•  Protocol converter (gateways)
•  OPC server
•  Common alarm via the relay outputs of the PRC487 control device
•  Conversion between BMS bus and digital inputs and outputs by means of alarm indica-

tor and operator panels (TM...) or signal converters (SMO480-12, SMO482-12, SMI 473-12.

Summary of Contents for MEDICS UMC710D4-160-HA

Page 1: ...Manual EN UMC710D4_D00379_00_M_XXEN 01 2020 MEDICS UMC710D4 HA Four pole switchover module for medical locations with and without manual automatic control...

Page 2: ...cents 30 sec remaining time 6 3 cents min Mobile phone higher depending on mobile phone tariff Repair service Repair calibration and replacement service Londorfer Strasse 65 35305 Gr nberg Germany Te...

Page 3: ...onality UMC710D4 11 2 5 1 The changeover module 11 2 5 2 Version UMC710D4 HA 12 2 5 3 Monitoring the device functions 13 2 5 4 Monitoring of the neutral conductor 13 2 5 5 Power supply 13 2 6 System c...

Page 4: ...23 6 Periodic verification and service 25 6 1 Periodic verification 25 6 1 1 Testing of the changeover module 26 6 2 Maintenance 26 7 Data manufacturer s certificate checklist circuit documentation 27...

Page 5: ...risk of danger that can lead to death or serious injury if not avoided I I Caution Indicates a low level risk that can result in minor or moderate injury or damage to property if not avoided i i Info...

Page 6: ...recommended by Bender Catastrophes caused by external influences and force majeure Mounting and installation with device combinations not recommended by the manufac turer This operating manual and the...

Page 7: ...4_D00379_00_M_XXEN 01 2020 7 MEDICS UMC710D4 HA Before installing and connecting the device make sure that the installation has been de ener gised The rules for working on electrical systems must be o...

Page 8: ...8 UMC710D4_D00379_00_M_XXEN 01 2020 System describtion...

Page 9: ...on has been cleared with Bender in advance Please note the limits of the area of application indicated in the technical data Use deviating from or beyond the scope of this is considered non compliant...

Page 10: ...or TN S systems SMI472 Signal converter for third party technical equipment e g med gases UPS TM Alarm indicator and operator panel UFA107E Changeover and monitoring module for IT systems with EDS ins...

Page 11: ...changeover module line 3 Monitoring of the switching elements to ensure that they are in the correct switching state and for wire breaks connecting wires respectively the circuits of the drive motors...

Page 12: ...ation in operating theatres or other essential sources of light etc For changeover modules with Ie 100 A the idle time is factory set to at least 500 ms Therefore the factory set total changeover time...

Page 13: ...eration generally small currents of at least 10 mA are assumed to flow continuously through the neutral conductor The connection of an additional load avoids the likelihood of completely symmetrical l...

Page 14: ...Fuses F1 F3 3 X3 Terminal strip X3 4 X5 Terminal strip X5 5 6N1 Control device PRC487 6 6T1 Power supply unit AN450 for the supply of up to 2 alarm indicator and test combinations MK2430 7 7N1 Relay m...

Page 15: ...when the anticipated short circuit current occur the protective device upstream of a fault will selectively trip the protective devices that are further upstream Section 710 537 6 2 VDE section 710 4...

Page 16: ...for the IT sys tem transformer must therefore be at least two stages smaller in rating than the fuse calculated using the formula above for the changeover module Of course the total nominal current fo...

Page 17: ...disconnected one open branch is connected to terminals X3 29 and X3 30 while the other one is con nected to terminals X3 31 and X3 32 The 120 resistor provided ex works must be removed and the open bu...

Page 18: ...18 UMC710D4_D00379_00_M_XXEN 01 2020 Commissioning setting and testing...

Page 19: ...y for the changeover process of the automatic changeover module Setting Response time t off If several changeover modules are connected in series in a power supply system it is re commended that they...

Page 20: ...Alarm adress 4 2 Second MK Adress 2 Alarm adress 4 1 Two changeover and monitoring modules with two alarm indicator and test combinations each Ger t Parameter Address setting for the first changeover...

Page 21: ...been changed to manual mode Automatic changeover will no longer be performed Short circuit distribution board Short circuit downstream of changeover module Following a failure on line 1 and changeover...

Page 22: ...is more than one master in the network More than one slave has the same address Interface A B polarity is incorrect No MASTER There is no master No alarm indicator and test combination or operator pa...

Page 23: ...ally in the event of a complete failure of the changeover module For that purpose proceed as follows 1 Remove the covers of the control section 2 Open the 3 pole fuse disconnectors F1 and F2 3 Remove...

Page 24: ...ctor Prerequisite Switch disconnector is switched off and charged displays 0 off and charged 4 Tension lever Charges the switch spring Press tension lever several times until charged is displayed If t...

Page 25: ...er switching device the IT system monitoring as well as the interaction of the components in the system The test encompasses the following Inspection Labels display elements mechanical components wiri...

Page 26: ...sage Failure line 2 2 Message on PRC487 K1on K2off Reconnect line 2 1 Alarm message Failure line 2 goes out 2 Message on PRC487 K1on K2off TEST on PRC487 Press the TEST button RESET button and MENU bu...

Page 27: ...ata manufacturer s certificate checklist circuit documentation 7 1 T V test report T V Technical Inspection Authority Service GmbH T V S d Gruppe Munich tested the change over module of the MEDICS ser...

Page 28: ...400 230 V Operating range Ue 0 8 1 15 x Ue Frequency fe 50 60 Hz Rated operational current Ie acc to DIN VDE 0100 710 see ordering information Fuse UMC710D4 29 see ordering information Utilisation cat...

Page 29: ...onnection rigid flexible conductor sizes 6 35 mm2 AWG 8 2 from 63 A direct connection Stripping length 23 mm General data EMC immunity acc to EN 61000 6 2 EMC emission acc to EN 61000 6 4 Classificati...

Page 30: ...00 HA 2 2 500 300 130 2 x 2 4 2 x 500 600 280 400 66 UMC710D4 630 HA 2 2 500 300 130 2 x 2 4 2 x 500 600 280 400 66 7 4 Ordering information Type Switching compo nents used Max operating current of th...

Page 31: ...6 TMX HA PQ PEM575 with PEM575 B22030107 TMX HA L2 L1 cable 2 left cable 2 right B22030108 7 5 Manufacturer s certificate checklist circuit documentation The individual documents complied for your MED...

Page 32: ...e www bender de All rights reserved Reprinting and duplicating only with permission of the publisher Bender GmbH Co KG PO Box 1161 35301 Gruenberg Germany Londorfer Str 65 35305 Gruenberg Germany Tel...

Reviews: