Hyperthermia Pump: Parameters Setting and Preventive Maintenance
Page 3 of 20
A.
SYSTEM SETUP
Changes in system setup can be made to:
Date and time: Set the real time clock and date
Audible alarm volume: Set the audible alarm volume level
Display brightness: Change the display brightness
Key Rate: Set touch key sensitivity
Pressure limits for High Pressure alarm: Set the maximum allowable in-line pressure. The possible
setting range from 100 to 300 mmHg.
Parameter Setup changes is performed in the Service mode.
www.belmontMedTech.com
Pressing the SERVICE key accesses the Calibration/Set-up mode. This key appears on the BELMONT logo
screen only at system powered-up. This screen remains active for 4.5 seconds before the system enters the
Prime mode
.
For Hyperthermia Treatment
Only. Not for infusion into
the circulatory System.
AGREE
DIS-
AGREE
CALIBRATION/SET-UP
TIME 23:59 DATE: 08-29-08
PRESS LIMIT 300mmHg
AC POWER PRESENT
EXIT
SERVICE
DISPLAY
BRIGHT
HARD-
WARE
PRESS
LIMIT
TEMP
CAL
PRESS
CAL
POWER
CAL
MEDIUM
KEYRATE
DATE
TIME
ALARM
VOL SET
Rev
CKSUM
=
AC POWER PRESENT
SERVICE