21

new access code.   The Timer will display 

0000  

with

one digit flashing.  The flashing digit may be adjusted
using   the  

*

  or  

&

  push   button,   when   set   as

required operating the 

(

 button will transfer control

to   the   next   digit.       When   all   the   digits  have   been
entered   press  

)

  twice   to   return   to   the  

AC5Et t

prompt   in   the   configuration   menu.   The   revised
access code will be activated when the BA577E-SS
is returned to the display mode.

Please contact BEKA associates sales department if
the access code is lost.

5.5.16 Direction of count:   

uP or dn

The Timer may be configured to

 time-up

 from zero to

the set time  

5Et t

  while displaying elapsed time, or

to  

time-down

  from the set time  

5Et t

  to zero while

displaying the remaining time.

When   the   repeat   timing   cycle   function  

CYCLE5

  is

enabled,   it   is   recommended   that   a   down   count   is
selected   so   that   the   progress   of   the   timer  can   be
observed with a known completion time i.e zero.   If
set as an up counter, elapsed will be displayed, but

5et t

 at which the Timer will stop is not shown.

To check the direction of count, select 

uP or dn

 from

the   configuration   menu   and   press  

(

  which   will

reveal the existing setting which can be changed by
pressing  the  

*

  or  

&

  button  followed  by  the  

)

button   to   enter   the   selection   and   return   to   the
configuration menu.

5.5.17 Power Fail:   

P-FAiL

Defines   how   the   Timer   powers-up   and   functions
when   power   is   restored   after   a   power   supply
interruption.     Three   options   are   available,  

idle

,

Pau5e

 and 

continue. 

idle

The Timer is stopped in the state it achieves

when it has timed-up to 

5et t

 or timed-down

to 

0000

,

,

with the elapsed or remaining time

when power was lost shown on the upper
display.   The Timer must be reset before it
can be restarted.   If the repeat timing cycle
is in use the number of cycles completed
will be lost when the Timer is reset.   

Pau5e

The Timer is stopped in the state it achieves

following   receipt   of   a   stop   input   to   pause
timing.     The   elapsed   or   remaining   time
when power was lost is shown on the upper
display.   Timing   resumes   when   a   start
instruction is received.  If a start input exists
when   power   is   restored   timing   will   start
immediately.

Continue

When   power   is   restored   the  Timer   will

continue from where it stopped without any
manual intervention.

To check or change the function, select 

P-fail

 from

the   configuration   menu   and   press

  (

  which   will

reveal the existing setting which can be changed by
pressing   the

  *  

or

  &

  button   followed   by  the

  )

button   to   enter   the   selection   and   return   to   the
configuration menu.

5.5.18 Local reset:     

LoC r5Et

The Local reset function contains two separate sub-
functions    

rE5Et

 

.

 

enbl

  and  

clr Gtot

  which   when

enabled allow the Timer and the grand total to be
reset via the instrument's front panel push buttons
while the Timer is in the display mode.

5.5.19 Local total reset:   

r5Et

 

.

 

enbl

r5Et

 

.

 

enbl

 is a sub-function in the 

Loc r5Et

 function

which when activated allows an operator to reset the
Timer  from  the  display  mode   by  operating  the  

*

and  

&

  push buttons simultaneously for more than

three seconds.

To   check   or   change   the   local   total   reset   select

LoC r5Et

  in   the   configuration   menu   and   press

(

 which will reveal the 

r5Et

 

.

 

enbl

 prompt, press 

(

again to show if the local total reset is 

on

 or 

oFF

.   If

set as required operate the 

)

 button twice to return

to the configuration menu, or the 

*

 or 

&

 button to

change the setting followed by the 

)

 button twice to

enter the change and return to the 

LoC

 

.

 

r5Et

 prompt

in the configuration menu.

Note:

The   Timer   may   also   be   reset   remotely   by
connecting   terminals  RS1   and   RS2   together
for more than one second.  See section 2.3

5.5.20 Local grand total reset:   

Clr Gtot

The grand total is the total run-time of the Timer that
may be viewed by operating the  

)

  and  

*

  push

buttons simultaneously in the display mode.   When
activated   

clr Gtot

  allows an operator to reset the

grand total display to zero from the display mode by
operating   the  

)

  and  

*

  push   buttons

simultaneously for more than ten seconds.

Clr Gtot

 is a sub-function in the 

LoC r5Et

 menu.  To

check or change the setting select  

LoC r5Et

  in the

configuration menu and press  

(

  which will reveal

rE5Et

 

.

 

enbl

.   Using  the  

&

  or  

 button  select

Clr Gtot

 and press 

(

 which will show if local grand

total reset is  

on

  or  

oFF

.     If set as required operate

the  

)

  button   twice   to   return   to   the   configuration

menu, or the 

&

 or 

*

 button to change the setting

followed by the 

)

 button twice to enter the change

and   return   to   the  

LoC r5Et

  prompt   in   the

configuration menu.

Summary of Contents for BA577E-SS

Page 1: ...Issue 6 4th May 2021 BA577E SS Rugged one input general purpose Timer or Clock Issue 6...

Page 2: ...t 5 5 11 Repeat timing cycle CYCLE5 5 5 12 Cycle function enable EnbL 5 5 13 Cycle count cycl cnt 5 5 14 Restart delay r5t dela 5 5 15 Adjusting the set time 5Et t and restart delay r5t dela from the...

Page 3: ...4 8 Enable Control output 1 EnbL 8 4 9 On and off times oP 1 on 1 oP 1 off 1 oP 1 on 2 oP 1 off 2 8 4 10 Control output 2 Optional oP2 8 4 11 Enable control output 2 EnbL 8 4 12 On and off times oP2...

Page 4: ...circuit that is used for all BA577E SS Timer installations Clock installations are the same except that the input terminals 5 6 are not used For simplicity the optional control outputs are described s...

Page 5: ...etic pick off sensor which is not suitable for controlling the BA577E SS 2 2 5 Voltage input Two voltage input ranges are independently selectable in the BA577E SS Timer configuration menu VoLt5 L and...

Page 6: ...BEKA stainless steel support plate should be slid over the rear of the Counter before the panel clamps are fitted to evenly distribute the clamping force and prevent the enclosure panel being distort...

Page 7: ...g the requested units of measurement if this information is not supplied when the instrument is ordered a blank card will be fitted A pack of self adhesive scale cards printed with common units of flo...

Page 8: ...isplay through which a scale card showing the units of time such as hours can be seen New instruments are fitted with a scale card showing the units of measurement specified when the instrument was or...

Page 9: ...ished via the front panel push buttons or by a remote contact A grand total time is maintained which can be viewed by operating the front panel push buttons It may be reset to zero from within the con...

Page 10: ...5 Shows in succession firmware version number instrument function elap5e and any output accessories that are fitted A Dual Control Outputs Access to configuration menu 5 3 Displays when configured as...

Page 11: ...d of the delay time automatically returns to Running Complete Entered automatically from Running when there are no more timing cycles to perform When the CYCLE5 function is enabled the timing cycle ca...

Page 12: ...he instrument using the or push button to adjust the flashing digit and the push button to transfer control to the next digit If the correct code has been entered pressing will cause the first paramet...

Page 13: ...itch contact Energise input by linking terminals 3 4 See section 5 5 5 dEbounCE Defines the level of debounce applied to the input to prevent false functioning dEFAuLt HEAVY LiGHt See section 5 5 6 Di...

Page 14: ...e display mode to 5Et t and r5t dela See section 5 5 15 Display Summary of function uP or dn Direction of count Defines whether the Timer times down from 5Et t to zero or times up from zero to the set...

Page 15: ...control output 1 See section 5 5 26 Display Summary of function CLr Gtot Resets grand total to zero This function resets the grand total which represents the total Timer run time from within the conf...

Page 16: ...s required press twice to return to the configuration menu or repeatedly press the or button until the required type of input is displayed and then press twice to return to the configuration menu One...

Page 17: ...n at start of each period 02 11 Current Number of cycles cycle requested not shown number when cycle is continuously repeated LabEL Exactly the same as 5td but timer status i e CYCLE or dELAY is shown...

Page 18: ...18...

Page 19: ...19...

Page 20: ...l to the next digit When set as required enter the selection and return to the cycl cnt prompt in the sub menu by operating the button 5 5 14 Restart delay r5t dela This sub function defines the time...

Page 21: ...mediately Continue When power is restored the Timer will continue from where it stopped without any manual intervention To check or change the function select P fail from the configuration menu and pr...

Page 22: ...uration example in section 6 2 of this manual illustrates the use of the control outputs Further examples can be found in Application Guide AG374 which is available on the BEKA website 5 5 22 Control...

Page 23: ...it can be adjusted using the or button and the button to move to the next digit When the required delay has been entered press to return to the control output 1 sub menu The Timer configuration exampl...

Page 24: ...rol to the second digit which should be set to u When 5urE has been entered pressing the button will reset all the configuration functions and return the instrument to the display mode Note r5Et dEF d...

Page 25: ...he required instrument configuration for this example are shown below Function Display Setting Access code CodE 0000 Function FunCtion ELAP5E Input inP type ContACt debounce dEbounCE dEFAuLt Display 2...

Page 26: ...r configuration menu Incorrect security code That the correct security code is being used Contact BEKA if code is lost 7 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING MAINTEN...

Page 27: ...e regularly checked Inspection frequency should be chosen to suit the environmental conditions 7 5 Guarantee Instruments which fail within the guarantee period should be returned to BEKA associates or...

Page 28: ...rument s permanent memory Following initialisation the instrument will display a flashing display which is a request for the local time to be entered The clock will not start to function until a displ...

Page 29: ...hroughout this manual push buttons are shown as and and legends displayed by the Clock are shown in a seven segment font as they appear on the instrument e g di5PLAY and CodE Access to the configurati...

Page 30: ...30...

Page 31: ...ntrol output 1 Optional Control output 1 can turn on output closed and off output open twice in each twenty four hour period This function contains five sub function EnbL Enables output 1 and defines...

Page 32: ...et time is entered the Clock will display a flashing 00 00 00 or 00 00 for 24 hour formats or A12 00 00 or A12 00 for 12 hour formats To adjust the clock set time select 5Et from the configuration men...

Page 33: ...tered return to the configuration menu by pressing the button twice 8 4 10 Control output 2 Optional oP2 Control output 2 is an optional factory fitted output which can be configured to turn on output...

Page 34: ...e display mode may be protected by a four digit alphanumeric security code which must be entered to gain access Default security code 0000 allows unrestricted access With ACCd displayed press to enter...

Page 35: ...g will then change the instrument to a Clock and return the display to FunCtion in the configuration menu Step 3 Select display format Using the or button select di5PLAY in the configuration menu and...

Page 36: ...access to the configuration menu Using the and buttons select CodE from the configuration menu and press which will reveal 0000 with the first digit flashing This example requires the security code to...

Page 37: ...ANCE If a BA577E SS fails after it has been functioning correctly as a Clock the following table may help to identify the cause of the failure Symptom Cause Check No display No power supply That there...

Page 38: ...8 10 6 Customer comments BEKA is always pleased to receive comments from customers about our products and services All communications are acknowledged and whenever possible suggestions are implemented...

Reviews: