5.2 Root extractor:  root

This function is primarily intended for use with
differential flowmeters which have a square law
4/20mA output. To activate the square root
extractor select 'root' from the menu and press 

P

which will reveal if the function is 'On' or 'OFF'. If
the function is set as required, press 

E

 to return to

the menu, or press the 

Up

 or 

Down

 button to

change the setting, followed by the 

E

 button to

return to the main menu.

For reference, the following table shows the output
current from a non-linearised differential flowmeter.

% of full flow

Current output mA

2.5

4.01

10.0

4.16

25.0

5.00

50.0

8.00

75.0

13.00

100.0

20.00

5.3 Resolution:  rESn

This function defines the resolution of the least
significant display digit. Decreasing the display
resolution can improve the readability of a noisy
signal. Select 'rESn' from the menu and press 

P

which will reveal the current display resolution. To
change the resolution press the 

Up

 or 

Down

 button

to select 1, 2, 5 or 10 digits, followed by the 

E

button to return to the menu.

5.4 Position of the decimal point: d.P.

A dummy decimal point can be positioned between
any of the digits or may be absent. To position the
decimal point select 'd.P.' from the menu and press

P

 . The decimal point can then be moved or turned

off by pressing the 

Up

 or 

Down

 push-button,

followed by 

E

 to return to the menu.

5.5 Calibration using an external
      current source:  CAL

This function enables the zero and span of the
indicator to be adjusted using an external
calibrated current source. When used with an
accurate traceable current source this is the
preferred method of calibration.

To calibrate the indicator select 'CAL' from the
main menu and press P. The indicator will display
'ZErO' which is a request for a 4mA input current.
Adjust the external current calibrator to 4.000mA
and again press 

P

 which will reveal the current

zero display. Each digit of the indicator display can
be changed by pressing the 

Up

 or 

Down

 buttons.

When the first digit is correct pressing 

P

 will

transfer control to the next digit. When all the digits
have been adjusted, press 

E

 to enter the new zero.

The indicator will display 'Ent' for a few seconds
while the information is being stored in memory,
and will then return to the 'ZErO' prompt.

To adjust the display at 20mA, press the Up button
which will cause the indicator to display 'SPAn'.
Adjust the external current calibrator to 20.000mA
and again press 

P

 which will reveal the existing full

scale display. Each digit of the indicator display
can be changed by pressing the 

Up

 or 

Down

buttons. When the first digit is correct pressing 

P

will transfer control to the next digit. When all the
digits have been adjusted press 

E

 to enter the new

span. The indicator will display 'Ent' for a few
seconds while the information is being stored in
memory, and will then return to the 'SPAn' prompt.
Finally press 

E

 again to return to the main menu.

This function may also be used when calibrating a
complete loop from primary element to indicator
display e.g. when the indicator is displaying the
output from a resistance thermometer to current
converter.

5.6 Calibration using internal references:  SEt

This function enables the zero and span of the
indicator to be adjusted without the need for an
accurate external current source, or for
disconnection from the 4/20mA loop.
The accuracy of this method depends upon the
accuracy of the internal references which should
be regularly calibrated as described in section
5.7.2

To calibrate the indicator select 'SEt' from the main
menu and press 

P

. The indicator will display 'ZErO'

, pressing 

P

 again will reveal the current display at

4mA. Each digit of the indicator display can be
changed by pressing the 

Up

 or 

Down

 buttons.

When the first digit is correct pressing 

P

 will

transfer control to the next digit. When the least
significant digit has been adjusted, press 

E

 to

return to the 'ZErO' prompt which completes the
adjustment.

To adjust the display at 20mA, press the 

Up

 button

which will cause the indicator to display 'SPAn'.
pressing 

P

 again will reveal the indicator display.

Each digit can be changed by pressing the 

Up

 or

Down

 buttons. When the first digit is correct

pressing 

P

 will transfer control to the next digit.

When the least significant digit has been adjusted
press 

E

 to return to the 'SPAn' prompt followed by

E

 to return to the main menu.

Summary of Contents for BA527C

Page 1: ...Issue 3 5th March 2009 BA527C BA528C loop powered 4 digit panel mounting indicators Issue 3...

Page 2: ...urce 6 2 Using the internal calibrator 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comme...

Page 3: ...o the 4 20mA input current provides the input signal for the analogue to digital converter Fig 1 Simplified block diagram Each time a 4 20mA current is applied to the instrument initialisation is perf...

Page 4: ...any 4 20mA signal to provide a remote indication Fig 3 shows a typical application in which the output from a gas analyser drives a BA527C or BA528C indicator Again it is necessary to ensure that the...

Page 5: ...trument and tighten until the indicator is secure as shown in Fig 5 Four clips are required to achieve an IP65 between a BA528C and the instrument panel Fig 4B BA528C dimensions c Connect the panel wi...

Page 6: ...this manual push buttons are shown in italics e g P or Up push button and legends displayed by the indicator are shown within inverted commas e g CAL and ALr2 Access to the programme menu is obtained...

Page 7: ...0mA loop See section 5 6 Display Description of function Cond Indicator conditioning This function provides access to a sub menu enabling the internal references to be calibrated and the mains line fr...

Page 8: ...8...

Page 9: ...adjusted press E to enter the new zero The indicator will display Ent for a few seconds while the information is being stored in memory and will then return to the ZErO prompt To adjust the display at...

Page 10: ...alibrator to 20 000mA and again press P The indicator will display Ent when the 20mA reference has been updated and will then return to the 020A prompt Two operations of the E button will return the i...

Page 11: ...requency of the mains supply is 50Hz and the existing security code is 1209 In the operating mode the indicator is required to display the input current as a percentage of span when the P push button...

Page 12: ...ain menu Step 8 Return to the display mode Return to the display mode by pressing E 6 2 Using the internal calibrator As in 6 1 the BA527C is required to display 50 0 with a 4mA input 1050 0 with a 20...

Page 13: ...ing If a BA527C or BA528C fails to function during commissioning the following procedure should be followed Symptom Cause Solution No display Incorrect There should be 1V wiring between terminals 1 3...

Page 14: ...d 7 6 Customer comments BEKA associates is always pleased to receive comments from customers about our products and services All communications are acknowledged and whenever possible suggestions are i...

Page 15: ...application of alarms Summary of programmable alarm functions Display Description of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section...

Page 16: ...y pressing the Up or Down button followed by the E button to return to the alarm menu 8 3 7 Hysteresis HStr During programming hysteresis is shown in the units the indicator has been calibrated to dis...

Page 17: ...security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to...

Page 18: ...18 Fig 12 Alarm programme menu...

Page 19: ...these functions the indicator initially displays the current break point and the total number of break points being used e g 5 13 current total number of break point break points Add Add a break point...

Page 20: ...ak points Each subsequent operation of the P push button will introduce an additional break point When adding a break point to a calibrated indicator the insertion position can be selected using the U...

Page 21: ...21...

Page 22: ...tivated The tare function allows an operator to quickly zero the indicator display at any input current so that subsequent readings only show the change in input since the P push button was operated W...

Reviews: