BEINAT TS1008-V1 User Manual And Maintenance Download Page 5

Page 5

Description of the screens

Probe sgm595
sn 24ad12
date  41   18
DY 2190
cy 5
al  15
dr 12
range 20
ADC   - 127 - 145

2

1

lel 20%  

100  ppm

fault

gas  methane

 

LPG  ecc

3

PRINT

OFS         

64    70  72    80

4

5

Tar      175   

180    200

*            print

PAGE 1  

to scroll the page press “ENTER”

Displays:   “Probe” The type of probe or device
Displays:   “SN”

The serial number

Displays:   “DATE”

The week and year of manufacture

Displays:   “Dy”

the remaining days at the end of the operating cycle.

Displays:   “Cy”

How many times the probe has been turned on.

Displays:   “AL”

How many times is gone into alarm.

Displays:   “DR”

How many corrections have been done

Displays:   “RANGE” The operating range if at 20 or at 100% of LEL
Displays:    “ADC”         A scrolling to the normalized value, coming out of the gas cap,
modifying  these  data  compromises  the  functioning  and  is  reserved  for  authorized
technicians.

To see the next page, press “DOWN”

PAGE 2  

to scroll the page press “ENTER”

Displays the percentage of gas dispersed in the environment, can be expressed in:
% of LEL or in ppm. In case of failure appears only “FAULT”
Displays the type of gas for which the probe has been calibrated

To see the next page, press “DOWN”

PAGE  3

PRINT  to print, point the IR transmitter toward the printer and press “ENTER”

PAGE  4

Modification:  “OFFSET”  changes  the  deviation  of  the  value  of  zero  defined  during
calibration.

PAGE  5

Modification: “TAR” multiplicative factor depending on the type of gas.

Printer recommended

BEINAT S.r.l. recommend a printer with infrared receiving 8 bits (1 start, error 4) Infrared
940nm, 33 kHz

Every time you disconnect the tester from the probe appears “PRINT”.
The TS1008 has loaded in memory the data stored in the probe and asked to print them.
If you do not want to print, turn off the tester or connect it to another probe.

Summary of Contents for TS1008-V1

Page 1: ... of the cycle Check and view how many times it has been turned on Check and view how many times it has detectedan Alarm Check and view how many corrections have been made Check and display the operating range of LEL Check and display the operation of the device Issues the receipt proving the operation TS1008 V1 Self certifying Tester for GAS Detectors and Sensors Certify the test by printing the r...

Page 2: ...s new device is suitable to test all conventional probes to detect explosive and toxic gas manufactured by BEINAT Srl This device allows to read all of the data and the configuration of work that are in the memory of the probes also having supplied by the serial transmission it prints the ticket that confirms the testing data certifying your own work By the Tester TS1008 you can read all of the ev...

Page 3: ...the 4 screws located at the back of the instrument 3 Raise the cohesion 4 Remove the batteries 5 Insert the new batteries positioning them in the correct direction 6 Replace the cover and tighten the 4 screws 7 Turn on the instrument and continue in your work Use Batteries must never be exposed to temperatures above 40 C To ensure optimal durability use them at room temperature If used at both low...

Page 4: ...on 6 Checking the status of the battery Pressing the UP and DOWN buttons at the same time the message BAT appears on the display followed by three segments Segments full charged battery empty segments empty battery DISPLAY The display used on the TESTER is of the alphanumeric type and is used to display the following information 3 LEL unit of measurement for explosive gas ppm for the toxic gas the...

Page 5: ...ctioning and is reserved for authorized technicians To see the next page press DOWN PAGE 2 to scroll the page press ENTER Displays the percentage of gas dispersed in the environment can be expressed in of LEL or in ppm In case of failure appears only FAULT Displays the type of gas for which the probe has been calibrated To see the next page press DOWN PAGE 3 PRINT to print point the IR transmitter...

Page 6: ...rged if it is not recharge it If the device does not detect Check that the cables are well plugged in and make contact If the device does not print Check that the instrument is in line with the printer Check that the printer is turned on and is in line with the instrument In the event other problems arise directly contact a specialist and or authorised technician or your BEINAT S r l dealer Proble...

Page 7: ...inder within the percentages described for each type of equipment Ordinary and extraordinary maintenance and putting out of service at the end of the operating period guaranteed by the manufacturer must be carried out by authorized and or specialized personnel LEL 20 Senso r Analyzer TS1008 Edit Enter UP Down Accessories To perform the test and to obtain a self certification it is necessary to use...

Page 8: ...nd from its enforced suspension of use DISPOSAL OF OLD ELECTRICAL ELECTRONIC EQUIPMENT This symbol on the product or its packaging to indicates that this product shall not be treated as household waste Instead it shall be handed over to the applicable collection point for the recycling of electrical and electronic equipment such as for example sales points in case you buy a new and similar product...

Reviews: