Cleaning the Distillation Head
During distillation, splashes of the digestion mixtures
can deposit in the distillation head. Although the test
program 00 will at the same time rinse the distillation
head, you will need to take the glassware out of the
device in order to clean it thoroughly at regular intervals.
If need be, you can change tubings, non-return valves
and other wear parts on this occasion. A Maintenance
Set is available which includes the tubings and other
wear parts that are most exposed to wear and tear. It
comes with a detailed instruction manual that is cur-
rently available in English or German.
Caution: the tubing may still hold remain-
ders of chemicals (especially, caustic
soda). Liquid leaking out might burn you or
damage the device.
Rinse the tubing with distilled water before
changing it.
Glass can break and cause injury! In work-
ing with glass components, observe all
appropriate safety precautions.
Lay out a towel or some other soft cloth beside the
device.
Take the sample vessel and the titration vessel out of
the device.
Pull the end cap off the steam delivery tube.
Undo the screw connection
that connects the
steam delivery tube to the distribution head.
Undo the two screw connections
that introduce
the delivery tubes into the glassware distillation head.
Pull the delivery tubes out of the distillation head. The
right one (the steam delivery tube) is loose now; lay
it aside. Leave the left tube (the NaOH delivery tube
) hanging behind the glassware.
Screw the cooling-water connectors
off the
condenser.
Screw off the lower of the two mounting brackets
which support the glassware.
Secure the glassware with one hand against toppling,
and screw off the other mounting bracket
.
Remember there is still water in the cooler; it might
drip out when dismantling the glassware.
Grasp the distillation head with the left hand and the
condenser with the right, and pull the glassware out
of the device.
Lay the glassware on the cloth beside the device.
Undo the screwcap
that fixes the condenser inlet
tube to the distillation head.
Now you can clean the distillation head and the con-
denser separately. To assemble the glassware again,
just reverse the steps described here.