7.0 Technical
Data
(at 23° C % 5° C, max. 75 % rel. humidity)
Display:
4+2+2, LCD incl. functions and symbols
Range Selection:
automatic
Auto-Power-Off:
approx. 30 min.
Overload display:
OL
Measuring rate:
2 measurements/sec. (LCD)
Clamp opening:
ca. 30 mm (1/2 inch)
Overvoltage Categorie:
CAT II, 600 V / CAT III, 300 V
Pollution degree:
2
Height above MSL.:
up to 2000 m
Battery display:
at low battery
Power supply:
Batterie 2 x 1,5 V IEC LR6 (alkaline)
Current Consumtion:
approx. 10 mA
Operation temperature:
-10° C … 50° C at less then 85% rel. H.
Storage temperature:
-20° C … 60° C at less then 75% rel. H.
Dimension:
210 x 6235,6 x 35,6 mm
Weight:
approx. 200 g
Voltage AC 50/60 Hz
Range
Resolution
Accuracy
5 V...600 V
0,1 V
± 0,5 % ± 5 D
Current AC 50/60 Hz (mA and A); Crest Factor <4
Range
Resolution
Accuracy
0,50...60 mA
0,01 mA
± 0,5 % rdg. +5 D
60...99,99 mA
0,01 mA
± 0,5 % rdg. +50 D
100...600 mA
0,1 mA
± 0,5 % rdg. +5 D
0,050...3,00 A
0,001 A
± 0,5 % rdg. +5 D
3,00...30,00 A
0,01 A
± 0,5 % rdg. +5 D
30,00...50,00 A
0,01 A
± 1,0 % rdg. +5 D
Overvoltageprotection: 600 A AC
Harmonics in mA, A Range (%)
Spezifiziert ab 100mA 50/60Hz (mA Range) oder ab 1A 50/60Hz
(A Range)
Range
Resolution
Accuracy
1...10
0,1 %
± 1 % rdg. ± 1%
11...20
0,1 %
± 5 % rdg. ± 1%
21...50
0,1 %
± 15 % rdg. ± 1%
51...99
0,1 %
± 35 % rdg. ± 1%
14