Now, the length measurement of the cable type cal-
ibrated can be performed, as described in section
6.2.
The reference range accuracy values depend
on the calibration of the customer specific
cable type. To comply with the accuracy values
in accordance with the specified technical data,
the user must perform the calibration of the
cable type as described in the instruction man-
ual.
The material selection copper / aluminium can
only be set during calibration. Once the refer-
ence range has been saved it is not possible to
change the material.
6.3.2 Deletion of Reference Ranges
If the user requires to assign the reference ranges
to other cable types, the reference range must first
be deleted.
Free memory locations of the reference ranges are
indicated by the display ''100.0''. Already assigned
memory locations are indicated by the display ''- -
-''.
"
Set UNITEST Cable Length Meter 3000 to the
reference range to be deleted. ''- - -'' is indicat-
ed on the display.
"
Press the
key again. The message
''100.0'' is displayed.
"
If the alligator clamps are open, press
and
keys. An error measurement is gener-
ated and indicated by the message ''Error''.
The reference range has now been deleted.
6.4 Performance of Cross Section
Measurements
The UNITEST Cable Length Meter allows the easy
and precise definition of unknown cross sections.
Only the length of the cable to be measured is re-
quired.
"
Switch on UNITEST Cable Length Meter press-
ing the
key.
"
Set measurement range selection dial to
''[Â/mm2]'' position. The message ''100.0 m''
is indicated on the display.
"
Set the material (copper / aluminium) of the
cable or conductor to be measured. The mes-
sage ''Cu'' or ''Al'' is indicated on the display.
"
The measurement unit (m / ft) can be selected
by pressing
and
keys.
"
Select the cable or conductor length by press-
ing
and ''
'' keys.
"
Connect the red test leads to red sockets and the
black test leads to black sockets.
"
Connect both conductor ends using the alliga-
tor clamps.
"
Press the
key to start the measurement.
A temperature compensation has been carried
out prior to performing the measurement. If an
external or manual temperature compensation
has to be carried out, please proceed as de-
scribed in section 6.1.
The conductor cross section is indicated on the dis-
play.
Shift
Shift
Shift
Shift
26
Performance of Measurements
m
Cu
T
CAL
TAB
Summary of Contents for Unitest 2005
Page 17: ...UNITEST Cable Length Meter 3000 Instruction Manual Cat No 2005 ...
Page 33: ...UNITEST Câblemètre 3000 Mode d emploi Réf 2005 ...
Page 49: ...UNITEST Medidor de cables 3000 Manual de instrucciones no de ref art 2005 ...
Page 66: ...66 ...
Page 68: ...68 Änderungen vorbehalten 03 2001 PTDB2005 00 Reg No 3335 ...