SPECIFICATIONS
Voltage Test (V) TRMS
Voltage range
1…1000 V AC (15…800 Hz),
1…1500 V DC(±)
LED nominal
voltage
+120 / -120 / 230 V: yellow LEDs
400 / 690 / 1000 V: red LEDs
LED tolerances
EN 61243-3
ELV indication LED
>50 V AC, >120 V DC: red LED
Impedance at ELV
level
320 kΩ @ 50 V AC
LED response time
<1s at 100% of each nominal value
LCD voltage range
1.0…1000 V, AC (15…800 Hz),
1.0…1500 V DC(±)
LCD resolution
0.1 V (1…29.9 V), 1 V (30…1500 V)
LCD accuracy
±(3% +15dgt) (1…29.9 V)
±(3% +3dgt) (30…1500 V)
Crest Factor
1 ... 330 V:
CF3
>330 ... 500 V:
CF2
>500 ... 1000 V:
CF1.41
Measurement duty
cycle
30 s ON, 240 s OFF
Acoustic indication
≥50 V AC, ≥120 V DC
Vibration
≥50 V AC, ≥120 V DC
Automatic power
ON
LED/LCD: > approx. 6 V AC/DC
Safety current
I <3.5 mA at 1000 V AC,
<6 mA at 1500 V DC
Single-Pole Phase Test (Pol)
Voltage range
100…1000 VAC against earth
Frequency range
50/60 Hz
Acoustic indication
Yes
Indication
same LED as ELV
Continuity Test (Rx)
Range
0…500 kΩ
Tolerance
0% to +50%
Acoustic indication
Yes
Indication
Yellow LED
Phase Rotation Indication
Voltage range
170…1000 VAC phase to phase
Frequency range
40…70 Hz
Indication
Green LEDs
Current Measurement (A) TRMS
Current range
0.1…200.0 A AC
Resolution
0.1 A
Tolerance
±(3% +5dgt)
Crest Factor
0 ... 100 A:
CF2
>100 ... 200 A:
CF1.41
Frequency range
50/60 Hz
Max. conductor size
13 mm
Resistance Measurement ( Ω )
Resistance range
0…100 kΩ
Resolution
1 Ω (1...2000 Ω), 1 kΩ (2...100 kΩ)
Tolerance
±(5% +10dgt) @ 25°C;
Acoustic indication
Sound < 20Ω
Non Contact Cable Break Detector / EF (Electric Field)
Range
100...1000 V AC (50/60 Hz)
Voltage Mode <10V
Specification
See voltage test
Frequency Measurement (Hz)
Frequency range
1…800 Hz