Attach D-‐Piece to monitor and Turn
on Gas Flow
Unit Calibrating
Calibration Successful
Calibration Failed
Retry Calibration
Firmware Version
6 Month Calibration Reminder
N/A
= Not Available on this device
Warnings and Maintenance
1.
Mouthpieces should be replaced after every use.
2.
Hands should be washed regularly in accordance with infection control practice.
3.
Please do not attempt to modify the equipment in any way or use accessories not specified
by the manufacturer. Any attempt to do so will invalidate the warranty and may
compromise the safety of the device.
4.
Bedfont will make available upon request service training to appropriately qualified persons.
5.
Holding the reset button down for 30 seconds will perform complete device reset,
this will
clear any saved data and revert all settings to the factory defaults
. After performing a reset
the device will need to have the date/time set and be calibrated before it can be used.
Cleaning
1.
The Smokerlyzer® products are moulded with SteriTouch® technology for optimum infection
control and bacterial efficacy. However Bedfont recommend wiping the instrument and D-‐
piece external surfaces with a product specifically developed for this purpose. Bedfont
provides instrument cleaning wipes. The D-‐Piece cannot be sterilised.
2.
NEVER use alcohol or cleaning agents containing alcohol or other organic solvents as long
term exposure to these vapours will damage the CO sensor inside.
3.
Under no circumstances should the instrument be immersed in liquid or splashed with
liquid.
Description
Micro+
basic
™
Attach Calibration Adaptor to D-‐Piece