Routine  Maintenance  

  

1.

 

Replace  batteries  when  indicated  by  the  empty  symbol  

.     

2.

 

Bedfont   recommend   removal   of   the   batteries   when   the   device   is   not   used   for   prolonged  
periods  of  time  to  prevent  leakage.  

3.

 

Replace   breath   sampling   D-­‐piece   every   30   days   or   if   visibly   soiled   or   contaminated.   The  

4.

 

The  sensor  should  be  replaced  every  5  years,  60  days  prior  to  the  sensor  change  ‘Countdown  
to   sensor   change’   symbol   will   be   shown   with   the   date   on   which   the   sensor   should   be  

changed.  This  can  be  ignored  by  pressing  

  until  the  date  at  which  the  sensor  should  be  

changed   arrives,   at   this   point   the   ‘Change   Sensor’   symbol   will   be   shown.   Change   sensor   if  
trained  to  by  an  approved  Bedfont  engineer  or  send  to  Bedfont  or  the  local  representative.  

5.

 

Additional  technical  information  can  be  made  available  on  request;  please  contact  Bedfont  

press  the  

  and  follow  instructions  to  calibrate  the  device  or  press  

  to  ignore.  

  

Troubleshooting  

  

The  unit  fails  to  turn  on:  

If  the  unit  fails  to  turn  on,  replace  the  batteries.  Ensure  that  the  batteries  are  inserted  the  correct  
way  around,  matching  the  symbols  moulded  into  the  plastic.  
  

High  readings  recorded  for  reported  non-­‐smoker:  

If   a   breath   test   on   a   non-­‐smoker   gives   an   indication   of   CO   being   present,   it   may   indicate   the  
following:  
  

1.

 

A  high  ambient  level  of  CO  (possibly  due  to  a  faulty  combustion  appliance  or  exhaust  fumes).  

2.

 

The  effects  of  passive  smoking.  

3.

 

The  device  is  contaminated  by  alcohol  or  another  organic  substance.  

4.

 

The  patient  may  have  a  very  high  level  of  hydrogen  on  their  breath,  which  can  be  generated  
in  the  human  digestive  system.  

1.

 

Ensure  the  fine  control  valve  is  in  the  off  position.    
  

2.

 

Screw  the  fine  control  valve  and  flow  indicator  assembly  to  the  gas  can.  This  is  best  done  by  
screwing  the  gas  can  into  the  valve.    

3.

 

Press  ‘Breath  Test’  symbol  on  screen,  

or

,  to  start  a  breath  test.    

5.

 

The  sensor  may  have  drifted  out  of  specification  (see  below).  

  

The  sensor  has  drifted  out  of  specification:  

Smokerlyzer®   products   are   calibrated   before   leaving  

coVita

,   however   we   recommend   if   you  

suspect   the   instrument   is   reading   incorrectly,   try   the   test   again   with   another   device   if   available  
to  get   a   comparison.   Alternatively   you   check   the   function   using  

coVita

   check   gas   or   send   back  

to  

coVita

.  The  check  gas  required  is  

coVita

  20ppm  carbon  monoxide  in  air,  procedure  below.  

    

Micro

+

basic

™   Smokerlyzer®   will   give   a   reminder   during   start-­‐up   when  

the   D-­‐piece   should   be   replaced,  see  ‘Change  D-­‐Piece’  symbol.  

or  its  distributor.  

6.

 

The   Micro

+

basic

™   Smokerlyzer®   requires   calibration   every   6   months   this   will   be   indicated  

by  

Summary of Contents for coVita Micro+basic Smokerlyzer

Page 1: ...Smokerlyzer Scientific contributions to health 0086 Former piCO For use with Micro basic www covita net 800 707 5751 ...

Page 2: ...d Device Symbols 4 Environment 5 Instrument Layout 5 Display and Device Symbols 6 Warnings and Maintenance 8 Cleaning 8 Routine Maintenance 9 Troubleshooting 9 Calibration Procedure 10 COdata Software connecting to the PC 10 Returns Procedure 11 Spares 11 Warranty 11 2 Converting COHb to FCOHb 3 Reviewing history 3 Technical Specification ...

Page 3: ...rbon monoxide is measured in parts per million ppm and blood carboxyhaemoglobin is measured in percentages COHb and fCOHb The three are compatible and convertible the ppm CO reading relates to gas in the lungs and on the breath i e the amount of poisonous CO that has been inhaled and the COHb and fCOHb readings relate to the percentage of vital oxygen that has been replaced in the bloodstream The ...

Page 4: ...conds unit will also power off after 2 minutes of inactivity to save power and then the history will then be shown Changing date and time Press and then you can then select either d m y or m d y for the date format and 12h or 24h for the time format Dates and times are then adjusted by selecting the number you wish to change and pressing or to increase and decrease Press to save your settings Conv...

Page 5: ...t Storage Pressure 800 1200mbar Operating Humidity 15 90 non condensing storage 0 95 Sensor Operating Life 5 years 5 year warranty Sensor Sensitivity 1ppm Sensor Drift 5 per annum Dimensions Approx 34 x 75 x 140 mm Weight Approx 200g including batteries Monitor Construction Case Polycarbonate ABS blend Micro basic Concentration Range CO 0 150ppm Display Full color touchscreen Detection Principle E...

Page 6: ...ith COdata Software 7 Single use SteriBreath mouthpiece 8 Battery compartment 9 Battery compartment clip 10 Reset button Environment The Micro basic Smokerlyzer complies with the directive EN60601 1 2 electromagnetic compatibility but can be affected by cellular phones and by electromagnetic interference exceeding the levels specified in EN50082 1 This equipment should be moved if n ecessary to av...

Page 7: ...pty Breath Test Environmental Test Settings Patient Profiles View patient results Inhale Hold Breath Countdown Timer Exhale ppm Reading COHb Reading Show fCOHb fCOHb Reading Display and Device Symbols Description Micro basic Battery condition Full ...

Page 8: ...Traffic Light Cut off Levels Save Change Breath Hold Time History Mute Change date time Next Step Countdown to Sensor Change Change Sensor Calibrate Device Attach Flow Meter to Gas Canister Description Micro basic Ambient CO reading ...

Page 9: ...own for 30 seconds will perform complete device reset this will clear any saved data and revert all settings to the factory defaults After performing a reset the device will need to have the date time set and be calibrated before it can be used Cleaning 1 The Smokerlyzer products are moulded with SteriTouch technology for optimum infection control and bacterial efficacy However Bedfont recommend w...

Page 10: ...t on a non smoker gives an indication of CO being present it may indicate the following 1 A high ambient level of CO possibly due to a faulty combustion appliance or exhaust fumes 2 The effects of passive smoking 3 The device is contaminated by alcohol or another organic substance 4 The patient may have a very high level of hydrogen on their breath which can be generated in the human digestive sys...

Page 11: ... lu b e h t e t a r lib a c o t w lo o o t is r o it n o m e h t f o e r u t a r e p m e t e h t If 4 r e i l p p su l ca o l r u yo r o t n o f d e B ct a t n co e s a le p s t is s r e p m le b o r p is h t if p u d e m ar w as h t i n u e h t e c n o r e at l n ai ag y Tr n w o sh e ill b w r e t e m o m r e h t d e r e h t e t a r lib a c o t h ig h o o t is r o it n o m e h t f o e r u t a r ...

Page 12: ...or loss or damage in transit When your goods are received you will be se nt an email stating so 6 If you choose not to proceed with the repair a handling fee will be charged Ensure that you return the completed authorization form with an Official Purchase Order Number 7 The equipment will be returned to you as soon as coVita has received all relevant paperwork A carriage fee will be charged if the...

Page 13: ...net 800 707 5751 coVita Tel 800 707 5751 Fax 800 721 2377 Email service covita net www covita net 212 Cottage Grove Ave Suite C Santa Barbara CA 93101 Issue 3 March 2017 Part No LAB679_USA Bedfont Scientific Limited reserves the right to change or update this literature without prior notice Registered in England and Wales Registered No 1289798 ...

Reviews: