Monthly Maintenance
Replace Alkaline Buffer
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-27
10
5
Separate the electrode (3), retainer (2) and retainer nut (1). (Refer to
Note: It is not required to clean the stir bar, however, the stir bar may be
cleaned if desired while the BUNm electrode is removed from the cup.
Refer to Four Month Maintenance, "
in this chapter.
6
Remove the quad-ring (4) from the front face of the electrode.
CAUTION
Avoid scratching gold coating on face of electrode as excessive
scratching may adversely affect performance.
Figure 10.14
7
Rinse the quad-ring in deionized water. Dry using lintless tissue.