Four-Month Maintenance
Cleaning MC Reagent Lines
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-50
October 2005
Four-Month Maintenance
Cleaning MC Reagent Lines
The MC Reagent Lines, Cups, and Stir Bars Cleaning procedure should be performed
every four months for ALBm, BUNm/UREAm, CREm, GLUCm
*
, PHOSm, and
TPm. More frequent cleaning may be required if erratic results, low results, or reagent
cups not filling are experienced.
Use the MC Maintenance Kit (P/N 474198) (includes 3 maintenance bottles with caps
and labels, and a bottle holder) to perform the cleaning.
Prepare the following before beginning this procedure:
• Two of the three maintenance bottles contained in the MC Line Maintenance kit are
to be filled with one of the prepared cleaning solutions. The third maintenance bottle
is filled with deionized water. Self-applied labels are included in the kit to label the
contents of each bottle.
• The maintenance bottle caps contain holes or openings into which the reagent straws
from the removed reagent bottles are inserted.
• The maintenance bottles are placed into a bottle holder and the holder is then
positioned onto the bottom frame of the MC reagent compartment. The back of the
holder hooks onto the bottom metal frame of the MC reagent compartment, so that
the holder sits upright on or near the floor.
*
Only the GLUCm cup module is used on DxC 600 Systems.
NOTICE
The system must be in
Standby
in order to access the
Maintenance
mode.
Table 10.2
MC Line Cleaning Preparation
MC Module
Prepare
By Diluting
ALBm, BUNm/UREAm,
GLUCm
a
, PHOSm
a
Only the GLUCm cup module is used on DxC 600 Systems.
10% cleaning solution
One part Wash Concentrate II
with nine parts deionized
water.
CREm, TPm
1N HCL
One part 6N HCL with five
parts deionized water.