Maintenance
Monthly Maintenance
9
1
Carefully replace MC cover over reaction cup modules and tighten screws.
IMPORTANT
While the ISE cover is removed, for maximum efficiency, the
Alkaline Buffer Damper Assembly
procedure (in this chapter) can be performed.
2
Reinstall the ISE module cover.
3
Calibrate all applicable cup chemistries.
Clean Mixers
Wipe the mixers monthly to clean any buildup or discoloration that may accumulate over time.
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
Gently but thoroughly wipe the outside of both mixers (1) and (2) of
using 70% isopropyl
alcohol and lintless tissue or with a 70% isopropyl alcohol swab. Inspect mixers for scratches or
nicks and replace if necessary.
Figure 9.17
CC Sample and Reagent Mixers
1.
CC Sample Mixer
2.
CC Reagent Mixer
A015939P.EPS
2
1