Quality Control
Daily QC
7
Daily QC
Daily QC consists of:
•
•
Running the AutoSetup Scheduler
•
Running the AutoSetup II Wizard
CAUTION
Risk of erroneous results if the Cytometer has been idle for an extended period of
time or you have just performed Daily Startup. To ensure correct results, perform
a prime after:
• Daily Startup.
• The Cytometer has been idle for an extended period of time.
• You place a new carousel on the MCL and light scatter signals appear
abnormal.
Preparing AutoSetup Samples
1
Prepare the Flow-Check Pro fluorospheres Follow the package insert instructions for mixing
and handling the fluorospheres.
2
Prepare the Flow-Set Pro fluorospheres. The Flow-Set Pro fluorospheres tube is used to set the
detector gains and voltages to the required level. Follow the package insert instructions for
mixing and handling the fluorospheres.
QC Process
QC Material Used
Verify fluidics and laser
alignment
Flow-Check Pro Fluorospheres. Verify HPCV versus expected value.
to the QC Database and review the QC data.
Adjust high voltage and
gain for a given application
Flow-Set Pro Fluorospheres. Ascertain target mean position based upon
application and adjust high voltage and gain daily to that target.
results to the QC Database and
.
Perform absolute counts
Flow-Count Fluorospheres.
Adjust color compensation
for a given application
For AutoSetup applications, Cyto-Comp Cells or whole blood stained with
QuickCOMP 2 or QuickCOMP 4 kit. Use single color stained samples with
each fluorochrome used in your application.
Verify correct settings with
an application Control
Update the control protocol with the settings derived from above. Run a
biological control equivalent to the application, such as Immuno-Trol Cells,
Immuno-Trol Low Cells, Cyto-trol Control Cells, or a normal whole blood.
to the QC Database and review the QC data.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......