773232AH
7-7
Quality Control
Autosetup II Wizard
7
Autosetup II Wizard
AutoSetup II simplifies and automates application setup and QC monitoring for multiple
applications simultaneously. You can perform a QC setup for all of your multicolor applications that
you run during a single day or shift at the same time using the
and
CAUTION
Risk of erroneous results if the Cytometer has been idle for an extended period of
time or you have just performed Daily Startup. To ensure correct results, perform
a prime after:
• Daily Startup.
• The Cytometer has been idle for an extended period of time.
• You place a new carousel on the MCL and light scatter signals appear
abnormal.
Before Running the AutoSetup II Wizard
•
Check that the
procedure was performed.
•
Allow about 40 minutes to warm up the system.
•
Check the
message. Start this procedure when the Cytometer status message
Awaiting Sample
appears. Record any error messages on the electronic
•
Workspace Preferences - Acquisition Options
to print results.
•
Workspace Preferences - LMD File Name
to specify listmode file names.
•
Follow the directions in the Flow-Check Pro fluorospheres package insert for details on storage
and handling.
•
Establish Flow-Set Pro fluorospheres target channels for your laboratory and modify the default
application definitions if necessary (see
3, one of the following options:
). Follow the directions
in the Flow-Set Pro fluorospheres package insert.
•
that you need to use with AutoSetup applications.
•
Set up region statistics exporting
and assign QC products to regions.
•
If acquisition ends as a result of duration (not stop count) while running AutoSetup II, select
on the Cytometer Control toolbar to continue.
•
Do not save scheduled AutoSetup Applications as a Worklist. Applications are intended to be
"scheduled". Do not add any protocols or panels to Acquisition Manager that has scheduled
Applications or AutoSetup II protocols. Always select "New Worklist" after scheduled QC
Applications or Autosetup protocols are run.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......