Sample Modules
Micro Liquid Module
question (?) mark next to each tube slot. Find All Tubes advances the carrousel to find all tubes in
the carrousel.
8.2.4 APS Troubleshooting
Micro Liquid Module
The LS 13 320 Micro Liquid Module (MLM) measures size distributions of particles suspended in
liquids from 0.4 microns to 2000 microns. It provides reliable results for researchers, quality control
laboratories, product or process control departments, or anyone needing to measure particle size
distributions.
Problem
Possible Cause/Solution
APS icon is not displayed
Menus are not displayed
•
Cables may be improperly connected.
•
Check connections (refer to section 2.2 for more
information on electrical and communication
connections)
Time out error “no communication with module”
•
Power cable to ASP not plugged-in
Does not perform according to selected SOM
•
Wrong SOM type.
•
Make sure the SOM selected is for the ALM + APS
Time out error “Auto-Prep Station tube pusher
motor timed-out”
•
Tube may be stuck in disposal slot. Check the
disposal slot.
•
Bar code label causes tube to get stuck. Check bar
code label for looseness.
Time out error “Auto-Prep Station tilt arm motor
timed-out”
•
Arm may be jammed. Check that modules are
aligned and the tubing is not caught on the side
Time out error “Auto-Prep Station sonicator motor
timed-out”
•
Call service
Time out error “Auto-Prep Station carousel motor
timed-out”
•
Tube may be stuck in disposal slot. Check the
disposal slot.
•
Bar code label causes tube to get stuck.
•
Check bar code label for looseness.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......