Connect the power cord
A power supply unit is used to supply power to the instrument.
Remove the magnetic stirrer
The magnetic stirrer in the docking module can be removed if necessary.
1.
Pull the tab on the docking module away from the instrument to
remove the module.
2.
Remove the four screws on the bottom plate of the module. Separate
the bottom plate from the module enclosure.
3.
Remove the screw that holds the magnet to the motor shaft. Hold the
shaft with pliers so that the shaft does not move.
4.
Remove the magnet and install the screw. Hold the shaft with pliers
so that the shaft does not move.
5.
Put the bottom plate on the module enclosure and install the four
screws.
User interface
The buttons on the front panel allow the user to operate the syringe,
sample probe and stirrer. Refer to
Figure 6
,
Table 1
and
Table 2
.
Figure 6 Front panel
1
Syringe controls
4
Stop button
2
Stirrer speed (RPM)
5
Sample probe controls
3
Power switch
6
LED status indicators
Table 1 Button descriptions
Button Name
Description
Load
Sends the syringe plunger to the top or bottom position.
Clean
Flushes the system.
Back-flush Back-flushes the system when held for 3 seconds.
English
15
Summary of Contents for HIAC 9703+
Page 2: ...English 3 Deutsch 21 Italiano 39 Fran ais 57 Espa ol 75 2...
Page 9: ...English 9...
Page 10: ...10 English...
Page 12: ...12 English...
Page 13: ...English 13...
Page 27: ...Deutsch 27...
Page 28: ...28 Deutsch...
Page 30: ...30 Deutsch...
Page 31: ...Deutsch 31...
Page 45: ...Italiano 45...
Page 46: ...46 Italiano...
Page 48: ...48 Italiano...
Page 49: ...Italiano 49...
Page 63: ...Fran ais 63...
Page 64: ...64 Fran ais...
Page 66: ...66 Fran ais...
Page 67: ...Fran ais 67...
Page 81: ...Espa ol 81...
Page 82: ...82 Espa ol...
Page 84: ...84 Espa ol...
Page 85: ...Espa ol 85...
Page 93: ......