Using the Editors While Creating Methods
3-13
Biomek® 3000 Laboratory Automation Workstation Quick-Start Guide
3.6
Using Hardware Setup
While a Beckman Coulter Service Engineer normally installs and configures new
devices in
Hardware Setup
, it may be necessary for you to install, configure, and
remove other devices using
Hardware Setup
if you change the configuration of your
instrument.
You will also need to configure
Hardware
Setup
if you have a new tool that you
want to use with your Biomek 3000 instrument.
You will add a P200L Single-Tip Pipette Tool to the list of tools available. Adding a
tool in
Hardware Setup
makes it available to add to a tool rack in the
Instrument
Setup
step and use it in a method.
To add a new tool in
Hardware
Setup
:
1. Choose
Instrument>Hardware Setup
.
Hardware Setup
appears (Figure 3-
Figure 3-13. Hardware Setup
2. In the left pane, right-click
Tools
and choose
Add Device
from the menu that
appears. A list of all the different tool types appears.
Right-click
Tools
bring up this menu.